FireCADCalc

UV Collimated Beam Dose with Petri, Water and Divergence Factors

Average UV dose delivered to a stirred sample under a collimated beam from radiometer irradiance and time, with the best-estimate uncertainty of the dose.

Inputs

Examples: Illustrative bench test: 0.12 mW/cm2 for 25 s, 12 mm sample, absorbance 0.08/cm, lamp at 45 cmHighly absorbing water: a 0.5/cm, 5 mm sample

Accepted: 0.0001 to 100 mW/cm²
Accepted: 0.1 to 3,600 s
Accepted: 0 to 20 %
Accepted: 50 to 100 %
Accepted: 0.0001 to 10 1/cm
Accepted: 1 to 100 mm
Accepted: 0.05 to 2 m
Accepted: 0 to 50 %
Accepted: 0 to 60 s
Accepted: 0 to 20 %
Accepted: 0 to 1 1/cm
Accepted: 0 to 10 mm
Accepted: 0 to 100 mm

Results

Bolton and Linden's collimated-beam protocol: irradiance corrected for reflection, the Petri factor, absorbance through the depth (water factor) and beam divergence. Uncertainty by root-sum-square of each variable's effect.

Water factor (1 − 10^(−ad))/(2.303 a d)76.029%
Divergence factor L/(L + d)99.010%
Average UV dose7.9266mJ/cm²
Best-estimate uncertainty of the dose0.46455mJ/cm²
Relative uncertainty5.8606%
Conservative dose7.4621mJ/cm²

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: Illustrative bench test: 0.12 mW/cm2 for 25 s, 12 mm sample, absorbance 0.08/cm, lamp at 45 cm. The inputs, in the units selected above:

InputValue
Irradiance at the sample surface centre0.12000 mW/cm²
Exposure time25.000 s
Reflection at the water surface2.5000 %
Petri factor (radiometer map average)92.000 %
Absorbance coefficient at 254 nm (base 10)0.080000 1/cm
Sample depth12.000 mm
Lamp-to-sample-surface distance0.45000 m
Uncertainty of E_s5.0000 %
Uncertainty of t0.30000 s
Uncertainty of P_f2.0000 %
Uncertainty of a_2540.005000 1/cm
Uncertainty of d0.50000 mm
Uncertainty of L5.0000 mm

Applying the formula below gives:

ResultValue
Water factor (1 − 10^(−ad))/(2.303 a d)89.719 %
Divergence factor L/(L + d)97.403 %
Average UV dose2.3516 mJ/cm²
Best-estimate uncertainty of the dose0.13285 mJ/cm²
Relative uncertainty5.6492 %
Conservative dose2.2188 mJ/cm²

Open this example in the calculator and change any input.

Formula

\(D_{CB} = E_s\,t\,(1 - R)\,P_f\,\frac{1 - 10^{-a_{254}d}}{2.303\,a_{254}d}\,\frac{L}{L + d}\)

\(U_D = \left[\sum_n\left(\frac{\partial D}{\partial V_n}U_{V_n}\right)^2\right]^{1/2}\)

Variables

Symbol Meaning SI unit
\(E_s\) Irradiance at the sample surface centre W/m²
\(t\) Exposure time s
\(R\) Reflection at the water surface -
\(P_f\) Petri factor (radiometer map average) -
\(a_{254}\) Absorbance coefficient at 254 nm (base 10) 1/m
\(d\) Sample depth m
\(L\) Lamp-to-sample-surface distance m
\(\delta E_s\) Uncertainty of E_s -
\(\delta t\) Uncertainty of t s
\(\delta P_f\) Uncertainty of P_f -
\(\delta a\) Uncertainty of a_254 1/m
\(\delta d\) Uncertainty of d m
\(\delta L\) Uncertainty of L m
\(WF\) Water factor (1 − 10^(−ad))/(2.303 a d) -
\(DF\) Divergence factor L/(L + d) -
\(D_{CB}\) Average UV dose J/m²
\(U_D\) Best-estimate uncertainty of the dose J/m²
\(U_D/D\) Relative uncertainty -
\(D - U_D\) Conservative dose J/m²

Method and standards

Average fluence in a stirred, collimated-beam sample with first-order error propagation.

Where it is used

  • UV dose–response tests for reactor validation.
  • Biodosimetry with MS2 or other surrogates.

Common mistakes

  • Using the decadic absorbance as a natural-log coefficient.
  • Skipping the Petri factor.

Assumptions and limits

  • Stirred sample; collimated low-pressure lamp at 254 nm.

References

  • Collimated-beam dose D = E_s t (1 − R) P_f (1 − 10^(−ad))/(2.303 ad) × L/(L + d) with root-sum-square uncertainty

Questions and answers

Why the water factor?

The sample absorbs UV with depth, so the average irradiance through the stirred depth is below the surface value by (1 − 10^(−ad))/(2.303ad).

Which uncertainty dominates?

Usually the radiometer (about 5 %), then the Petri factor and the timing.

What do the defaults represent?

An illustrative bench test: 0.12 mW/cm² for 25 s, 12 mm of water with an absorbance of 0.08/cm (8 /m), lamp 45 cm above the sample.

Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Sources

Crittenden et al., MWH's Water Treatment: Principles and Design, 3rd ed. (Wiley, 2012), sec. 13-9, eqs. 13-63 to 13-66 and Example 13-11

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.