Two-Colour (Ratio) Pyrometer: True Temperature from Emissivity Ratio

True temperature from a two-colour pyrometer reading when the target is not grey, from the two wavelengths and the ratio of emissivities (Wien approximation).

Thermal power plantProcess steam and utilitiesSugar mill and cogenerationDistillery and ethanolRefinery and petrochemicalChemical and fertiliser

Inputs

Examples: 0.9/1.05 µm, 1,200 °C, ε ratio 1.050.75/0.95 µm, 900 °C, ε ratio 0.97

Results

Wien approximation (valid for λT below about 3,000 µm·K). Ratio pyrometers tolerate dust, smoke and partial targets as long as the emissivity ratio stays near 1.

True temperature1,155.1°C
Indication error44.949K

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: 0.9/1.05 µm, 1,200 °C, ε ratio 1.05. The inputs, in the units selected above:

InputValue
Ratio temperature indicated1,200.0 °C
Shorter wavelength0.90000 µm
Longer wavelength1.0500 µm
Emissivity ratio1.0500 -

Applying the formula below gives:

ResultValue
True temperature1,155.1 °C
Indication error44.949 K

Open this example in the calculator and change any input.

Formula

\(\frac1T = \frac{1}{T_r} + \frac{\ln(\varepsilon_1/\varepsilon_2)}{c_2\left(\frac{1}{\lambda_1} - \frac{1}{\lambda_2}\right)}, \quad c_2 = 0.014388\ \mathrm{m\,K}\)

Variables

Symbol Meaning SI unit
\(T_r\) Ratio temperature indicated K
\(\lambda_1\) Shorter wavelength m
\(\lambda_2\) Longer wavelength m
\(\varepsilon_1/\varepsilon_2\) Emissivity ratio -
\(T\) True temperature K
\(T_r - T\) Indication error K

Method and standards

Wien ratio pyrometry.

Where it is used

  • Furnace, kiln and molten-metal temperatures.
  • Correcting ratio readings on oxidised steel.

Common mistakes

  • Assuming any target is grey.
  • Using Planck-range wavelengths with the Wien form.

Assumptions and limits

  • Wien's approximation is within about 1 % of Planck's law for λT below about 3,000 µm·K; the page warns when the longer wavelength times the temperature exceeds that.

  • Wien approximation.

References

  • Ratio pyrometry with Wien's approximation: 1/T = 1/T_r + ln(ε₁/ε₂)/(c₂(1/λ₁ − 1/λ₂))

Questions and answers

Why use two colours?

If emissivity is the same at both wavelengths, it cancels from the ratio; dirty windows and smoke that dim both channels equally do not matter.

When does it go wrong?

When the target is not grey: metals and oxides often have ε₁/ε₂ of 1.02–1.1, shifting the reading tens of kelvin.

What do the defaults represent?

A 0.9/1.05 µm ratio pyrometer reading 1,200 °C on a target with ε₁/ε₂ = 1.05.

Method and article reviewed by FireCAD engineering on 13 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.