Worked example
Example: 10 k, beta 3 950 at 50 °C; 6 kΩ measured. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Resistance at the reference temperature | 10,000 Ω |
| Reference temperature | 25.000 °C |
| β constant (K) | 3,950.0 - |
| Temperature (to find R) | 50.000 °C |
| Measured resistance (to find T) | 6,000.0 Ω |
Applying the formula below gives:
| Result | Value |
|---|---|
| Resistance at T | 3,588.2 Ω |
| Temperature at the measured resistance | 36.957 °C |
| Sensitivity at T, (1/R) dR/dT = −β/T² (per kelvin) | -0.037826 1/K |
Open this example in the calculator and change any input.
Formula
\(R_T = R_0\, e^{\beta(1/T - 1/T_0)}, \quad \frac1T = \frac{1}{T_0} + \frac{1}{\beta}\ln\frac{R}{R_0}\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(R_0\) | Resistance at the reference temperature | Ω |
| \(T_0\) | Reference temperature | K |
| \(\beta\) | β constant (K) | - |
| \(T\) | Temperature (to find R) | K |
| \(R_m\) | Measured resistance (to find T) | Ω |
| \(R_T\) | Resistance at T | Ω |
| \(T_R\) | Temperature at the measured resistance | K |
| \(\alpha\) | Sensitivity at T, (1/R) dR/dT = −β/T² (per kelvin) | 1/K |
Method and standards
Two-parameter NTC model.
Where it is used
- HVAC and appliance sensors.
- Motor winding protection.
Common mistakes
- Using °C in the exponent.
- Using β outside its fitted range.
Assumptions and limits
- NTC thermistor.
References
- NTC β model R = R₀ exp[β (1/T − 1/T₀)]
Questions and answers
What is a 10 k, β 3950 thermistor at 50 °C?
3588 Ω.
What temperature is 6 kΩ?
37.0 °C.
Why so sensitive?
About −4 % per kelvin near room temperature, ten times an RTD.
Method and article reviewed by FireCAD engineering on 12 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.