Worked example
Example: 10 k NTC: 0, 25, 50 °C points; 5 kΩ. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Temperature 1 | 0 °C |
| Resistance at T₁ | 32,650 Ω |
| Temperature 2 | 25.000 °C |
| Resistance at T₂ | 10,000 Ω |
| Temperature 3 | 50.000 °C |
| Resistance at T₃ | 3,603.0 Ω |
| Measured resistance | 5,000.0 Ω |
Applying the formula below gives:
| Result | Value |
|---|---|
| Coefficient A | 0.001125 |
| Coefficient B | 2.3472e-04 |
| Coefficient C | 8.5631e-08 |
| Temperature at the measured resistance | 41.580 °C |
Open this example in the calculator and change any input.
Formula
\(\frac1T = A + B \ln R + C (\ln R)^3\)
With \(L_i = \ln R_i\), \(Y_i = 1/T_i\) and \(\gamma_2 = \frac{Y_2 - Y_1}{L_2 - L_1}\), \(\gamma_3 = \frac{Y_3 - Y_1}{L_3 - L_1}\):
\(C = \frac{\gamma_3 - \gamma_2}{(L_3 - L_2)(L_1 + L_2 + L_3)}, \quad B = \gamma_2 - C\left(L_1^2 + L_1L_2 + L_2^2\right), \quad A = Y_1 - \left(B + C L_1^2\right)L_1\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(T_1\) | Temperature 1 | K |
| \(R_1\) | Resistance at T₁ | Ω |
| \(T_2\) | Temperature 2 | K |
| \(R_2\) | Resistance at T₂ | Ω |
| \(T_3\) | Temperature 3 | K |
| \(R_3\) | Resistance at T₃ | Ω |
| \(R_m\) | Measured resistance | Ω |
| \(A\) | Coefficient A | - |
| \(B\) | Coefficient B | - |
| \(C\) | Coefficient C | - |
| \(T\) | Temperature at the measured resistance | K |
| \(R\) | Resistance in the equation | Ω |
| \(L_i, Y_i\) | ln R_i (R in Ω) and 1/T_i | -, 1/K |
| \(\gamma_2, \gamma_3\) | Slopes between points 1–2 and 1–3 | K⁻¹ |
Method and standards
Three linear equations in A, B and C.
Where it is used
- Thermistor calibration.
- Programming instruments with ABC coefficients.
Common mistakes
- Points too close together.
- Mixing °C and K.
Assumptions and limits
- NTC thermistor; resistances in ohms inside the logarithms.
- No unique fit exists when R₁R₂R₃ = 1 Ω³ (ln R₁ + ln R₂ + ln R₃ = 0); the page then asks for other points. Real thermistor points in the kΩ range are far from this.
References
- J.S. Steinhart and S.R. Hart
- Deep-Sea Research 15 (1968) 497
Questions and answers
Why three points?
The equation has three unknowns, A, B and C.
How accurate is it?
Typically within a few hundredths of a kelvin over 100 K.
Is the default a real thermistor?
Points typical of a 10 k NTC; use your own calibration data.
Method and article reviewed by FireCAD engineering on 12 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.