Worked example
Example: Pt100 at 100 °C, 1 mA, 2 mW/K. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Measuring current | 1.0000 mA |
| Sensor resistance at the temperature | 138.50 Ω |
| Self-heating coefficient (mW per K) | 2.0000 - |
Applying the formula below gives:
| Result | Value |
|---|---|
| Power in the sensor I² R | 0.13850 mW |
| Self-heating error (reads high) | 0.069250 K |
Open this example in the calculator and change any input.
Formula
\(P = I^2 R, \quad \Delta T = \frac{P}{k_{sh}}\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(I\) | Measuring current | A |
| \(R\) | Sensor resistance at the temperature | Ω |
| \(k_{sh}\) | Self-heating coefficient (mW per K) | - |
| \(P\) | Power in the sensor I² R | W |
| \(\Delta T\) | Self-heating error (reads high) | K |
Method and standards
Power balance at the sensor.
Where it is used
- Choosing excitation current.
- Precision temperature checks.
Common mistakes
- Coefficient for the wrong medium.
- High currents for better signal.
Assumptions and limits
- Steady state.
- k_sh is in mW per K. Some data sheets give the inverse, a self-heating error in K/mW (°C/mW): enter 1 divided by that value.
References
- ΔT = I² R / k (IEC 60751 self-heating definition)
Questions and answers
1 mA through a Pt100 at 100 °C with 2 mW/K?
0.1385 mW, error 0.069 K.
And at 5 mA?
25 times more: about 1.7 K.
Why does the medium matter?
Flowing liquid carries heat away; still air does not.
Method and article reviewed by FireCAD engineering on 12 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.