Worked example
Example: Illustrative 2.5 mm LVDT. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Sensitivity (mV per V of excitation per mm) | 45.000 - |
| Excitation voltage (RMS) | 3.0000 V |
| Core displacement from null | 1.8000 mm |
| Full-scale range (±) | 2.5000 mm |
| Measured output (for the inverse) | 0.15000 V |
| Non-linearity (% of full scale) | 0.25000 % |
Applying the formula below gives:
| Result | Value |
|---|---|
| Output at the displacement | 243.00 mV |
| Output at full scale | 337.50 mV |
| Displacement for the measured output | 1.1111 mm |
| Error band from non-linearity | 6.2500 µm |
Open this example in the calculator and change any input.
Formula
\(V_o = S\,V_e\,x, \quad x_m = \frac{V_m}{S\,V_e}\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(S\) | Sensitivity (mV per V of excitation per mm) | - |
| \(V_e\) | Excitation voltage (RMS) | V |
| \(x\) | Core displacement from null | m |
| \(x_{FS}\) | Full-scale range (±) | m |
| \(V_m\) | Measured output (for the inverse) | V |
| \(NL\) | Non-linearity (% of full scale) | - |
| \(V_o\) | Output at the displacement | V |
| \(V_{FS}\) | Output at full scale | V |
| \(x_m\) | Displacement for the measured output | m |
| \(\pm e\) | Error band from non-linearity | m |
Method and standards
LVDT output.
Where it is used
- Displacement, valve position and gauge heads.
- Setting up LVDT signal conditioners.
Common mistakes
- Using the output beyond the linear range.
- Ignoring the excitation voltage in the sensitivity.
Assumptions and limits
- Linear range; RMS voltages; ideal demodulation.
References
- Linear variable differential transformer output
- Lipták (ed.)
- Instrument Engineers' Handbook
- Vol. 1 Process Measurement and Analysis and Vol. 2 Process Control and Optimization
- 4th ed. (CRC Press, 2003-2006)
Questions and answers
How is the direction of motion found?
From the output's phase relative to the excitation, using a phase-sensitive demodulator.
Why quote sensitivity per volt of excitation?
Output scales with excitation, so the data sheet gives it per volt for any excitation used.
What do the defaults represent?
An illustrative ±2.5 mm LVDT of 45 mV/V/mm on 3 V excitation, core at 1.8 mm, 0.25 % non-linearity.