FireCADCalc

Repeatability: Mean, Standard Deviation and Standard Uncertainty

Mean, standard deviation, standard uncertainty of the mean and the 95 % interval of up to ten repeated readings (Type A evaluation of uncertainty).

Thermal power plantProcess steam and utilities

Inputs

Examples: Ten pressure readingsFirst five of themThree temperatures 85.2, 85.6, 85.1 °C

Results

The 95 % interval of the mean is about ±t u with Student's t for n − 1 degrees of freedom (2.26 for 10 readings).

Mean100.15
Standard deviation (n − 1)0.21731
Standard uncertainty of the mean s/√n0.068718
Coefficient of variation0.21698%
Range (largest − smallest)0.70000

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: Ten pressure readings. The inputs, in the units selected above:

InputValue
Reading 1100.20 -
Reading 2100.50 -
Reading 399.800 -
Reading 4100.10 -
Reading 5100.40 -
Reading 6100.00 -
Reading 799.900 -
Reading 8100.30 -
Reading 9100.20 -
Reading 10100.10 -
Number of readings to use (2 to 10)10.000 -

Applying the formula below gives:

ResultValue
Mean100.15
Standard deviation (n − 1)0.21731
Standard uncertainty of the mean s/√n0.068718
Coefficient of variation0.21698 %
Range (largest − smallest)0.70000

Open this example in the calculator and change any input.

Formula

\(\bar x = \frac{1}{n}\sum x_i, \quad s = \sqrt{\frac{\sum (x_i - \bar x)^2}{n - 1}}, \quad u = \frac{s}{\sqrt n}\)

Variables

Symbol Meaning SI unit
\(x_{1}\) Reading 1 -
\(x_{2}\) Reading 2 -
\(x_{3}\) Reading 3 -
\(x_{4}\) Reading 4 -
\(x_{5}\) Reading 5 -
\(x_{6}\) Reading 6 -
\(x_{7}\) Reading 7 -
\(x_{8}\) Reading 8 -
\(x_{9}\) Reading 9 -
\(x_{10}\) Reading 10 -
\(n\) Number of readings to use (2 to 10) -
\(\bar x\) Mean -
\(s\) Standard deviation (n − 1) -
\(u\) Standard uncertainty of the mean s/√n -
\(s/\bar x\) Coefficient of variation -
\(R\) Range (largest − smallest) -

Method and standards

Type A evaluation of the GUM.

Where it is used

  • Instrument repeatability tests.
  • Laboratory results.
  • Quality control.

Common mistakes

  • Using n instead of n − 1.
  • Mixing different conditions.

Assumptions and limits

  • Independent readings.

References

  • JCGM 100:2008 (GUM) §4.2
  • Type A evaluation

Questions and answers

What do the ten default readings give?

Mean 100.15, standard deviation 0.21, standard uncertainty of the mean 0.069.

Why n − 1?

The mean is taken from the same data, which uses up one degree of freedom.

What is the coefficient of variation?

The standard deviation as a share of the mean.

Method and article reviewed by FireCAD engineering on 12 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.