Worked example
Example: 10 kg on 1 cm², g = 9.7866. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Mass of weights and piston | 10.000 kg |
| Effective piston area | 1.0000 cm² |
| Local gravity | 9.7866 m/s² |
| Air density | 1.2000 kg/m³ |
| Density of the weights | 7,920.0 kg/m³ |
Applying the formula below gives:
| Result | Value |
|---|---|
| Generated pressure | 9.7851 bar |
| Pressure assuming 9.80665 m/s² and no buoyancy | 9.8066 bar |
| Error of the simple calculation | 0.22006 % |
Open this example in the calculator and change any input.
Formula
\(p = \frac{mg_l(1 - \rho_a/\rho_m)}{A_e}\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(m\) | Mass of weights and piston | kg |
| \(A_e\) | Effective piston area | m² |
| \(g_l\) | Local gravity, in m/s² | m/s² |
| \(\rho_a\) | Air density | kg/m³ |
| \(\rho_m\) | Density of the weights | kg/m³ |
| \(p\) | Generated pressure | Pa |
| \(p_{std}\) | Pressure assuming 9.80665 m/s² and no buoyancy | Pa |
| \(\Delta p/p\) | Error of the simple calculation | - |
Method and standards
Pressure balance.
Where it is used
- Pressure gauge and transmitter calibration.
- Laboratory accuracy checks.
Common mistakes
- Using standard gravity.
- Ignoring piston temperature.
Assumptions and limits
- Effective area at reference conditions: the makers' temperature and pressure-distortion coefficients for the piston are not applied.
- No fluid-head correction between the piston reference level and the device under test.
- Local gravity is entered in m/s² in every unit system.
References
- Pressure balance p = mg(1 − ρ_a/ρ_m)/A_e (calibration practice; OIML R 110 principle)
Questions and answers
Why does local gravity matter?
Weights press with mg; gravity varies about 0.5 % between the equator and poles and with altitude.
Why air buoyancy?
Air displaced by the weights pushes up with about 0.015 % of their weight for steel weights.
What do the defaults represent?
10 kg on a 1 cm² piston at a gravity typical of central India.
Method and article reviewed by FireCAD engineering on 13 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.