Worked example
Example: 0–100, reading 20, ±0.5 % of span and of reading. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Lower range value | 0 - |
| Upper range value | 100.00 - |
| Reading | 20.000 - |
| Accuracy as a share of span | 0.50000 % |
| Accuracy as a share of reading | 0.50000 % |
Applying the formula below gives:
| Result | Value |
|---|---|
| Error band by % of span (±) | 0.50000 |
| Error band by % of reading (±) | 0.10000 |
| Span accuracy as a share of the reading | 2.5000 % |
Open this example in the calculator and change any input.
Formula
\(e_s = a_s\,(x_{hi} - x_{lo}), \quad e_r = a_r\, |x|\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(x_{lo}\) | Lower range value | - |
| \(x_{hi}\) | Upper range value | - |
| \(x\) | Reading | - |
| \(a_s\) | Accuracy as a share of span | - |
| \(a_r\) | Accuracy as a share of reading | - |
| \(e_s\) | Error band by % of span (±) | - |
| \(e_r\) | Error band by % of reading (±) | - |
| \(e_s/x\) | Span accuracy as a share of the reading | - |
Method and standards
Definitions.
Where it is used
- Instrument selection and range choice.
- Comparing data sheets.
Common mistakes
- Treating % span as % reading.
- Ignoring ambient effects.
Assumptions and limits
- Accuracy statement only; at a zero reading the relative value is not shown.
References
- Definitions of accuracy classes (IEC 60770, ISA-51.1 terminology)
Questions and answers
±0.5 % of a 0–100 span at a reading of 20?
±0.5, which is ±2.5 % of the reading.
And ±0.5 % of reading?
±0.1 at 20.
Which is better?
% of reading, at the low end of the range.
Method and article reviewed by FireCAD engineering on 12 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.