FireCADCalc

Isokinetic Sampling: Nozzle Flow and Meter Reading

Sample flow through a nozzle for isokinetic dust sampling, converted to dry gas meter conditions, and the nozzle size for a target meter flow.

Thermal power plantSugar mill and cogenerationRefinery and petrochemicalChemical and fertiliser

Inputs

Examples: Boiler stack 15 m/sFast stack 25 m/s

Results

Isokinetic: gas enters the nozzle at the stack velocity, so particles are neither over- nor under-sampled.

Sample flow at stack conditions29.452L/min
Dry flow at the meter17.702L/min
Nozzle for the target meter flow5.3147mm

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: Boiler stack 15 m/s. The inputs, in the units selected above:

InputValue
Stack gas velocity at the point15.000 m/s
Nozzle diameter6.0000 mm
Stack gas temperature140.00 °C
Stack pressure (absolute)100.00 kPa
Gas meter temperature25.000 °C
Gas meter pressure (absolute)99.000 kPa
Stack moisture (removed before the meter)10.000 %
Target meter flow (for nozzle choice)15.000 L/min

Applying the formula below gives:

ResultValue
Sample flow at stack conditions25.447 L/min
Dry flow at the meter16.694 L/min
Nozzle for the target meter flow5.6874 mm

Open this example in the calculator and change any input.

Formula

\(\dot V_s = v_s\frac{\pi d_n^2}{4}, \quad \dot V_m = \dot V_s\frac{T_m}{T_s}\cdot\frac{p_s}{p_m}(1 - B_{ws})\)

Variables

Symbol Meaning SI unit
\(v_s\) Stack gas velocity at the point m/s
\(d_n\) Nozzle diameter m
\(T_s\) Stack gas temperature K
\(p_s\) Stack pressure (absolute) Pa
\(T_m\) Gas meter temperature K
\(p_m\) Gas meter pressure (absolute) Pa
\(B_{ws}\) Stack moisture (removed before the meter) -
\(\dot V_t\) Target meter flow (for nozzle choice) m³/s
\(\dot V_s\) Sample flow at stack conditions m³/s
\(\dot V_m\) Dry flow at the meter m³/s
\(d_t\) Nozzle for the target meter flow m

Method and standards

Continuity and ideal gas.

Where it is used

  • Stack dust monitoring.
  • ESP and bag filter performance tests.

Common mistakes

  • Using meter flow without temperature correction.
  • Ignoring moisture removal.

Assumptions and limits

  • Ideal gas.

References

  • Isokinetic condition and ideal-gas correction to meter conditions (EPA Method 5 principle; ISO 9096)

Questions and answers

15 m/s, 6 mm nozzle, 140 °C, 10 % moisture?

About 16.7 L/min dry at the meter.

Why isokinetic?

Too slow over-samples coarse dust, too fast under-samples it.

Changing velocity across the duct?

Re-set the flow at each traverse point.

Method and article reviewed by FireCAD engineering on 12 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.