Capillary (Ostwald, Ubbelohde) Viscometer Calculation

Viscometer constant from a reference liquid, and the kinematic and dynamic viscosity of a sample from its efflux time, with the kinetic energy correction.

Inputs

Examples: Illustrative oil, water-calibratedShort-time tube with KE correctionViscous oil in a large-bore tube

Accepted: 0.1 to 100,000 cSt
Accepted: 1 to 100,000 s
Accepted: 1 to 100,000 s
Accepted: 100 to 3,000 kg/m³
Accepted: 0 to 1,000

Results

Gravity-driven capillary flow is Hagen–Poiseuille, so ν = Ct − B/t: the first term is laminar flow, the second corrects for the kinetic energy given to the liquid. C comes from a reference liquid (water is 1.0034 mm²/s at 20 °C) or the maker's certificate; B is usually negligible for efflux times above about 200 s. Dynamic viscosity μ = ρν. Keep the bath temperature within ±0.02 K for accurate work.

Viscometer constant (mm²/s²)0.002951
Sample kinematic viscosity1.5110cSt
Sample dynamic viscosity1.3297mPa·s
Kinetic energy correction share0%

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: Illustrative oil, water-calibrated. The inputs, in the units selected above:

InputValue
Reference liquid kinematic viscosity1.0034 cSt
Reference efflux time340.00 s
Sample efflux time512.00 s
Sample density880.00 kg/m³
Kinetic energy correction constant (mm²)0 -

Applying the formula below gives:

ResultValue
Viscometer constant (mm²/s²)0.002951
Sample kinematic viscosity1.5110 cSt
Sample dynamic viscosity1.3297 mPa·s
Kinetic energy correction share0 %

Open this example in the calculator and change any input.

Formula

\(C = \frac{\nu_{ref} + B/t_{ref}}{t_{ref}}, \quad \nu = Ct - \frac Bt, \quad \mu = \rho\nu\)

Variables

Symbol Meaning SI unit
\(\nu_{ref}\) Reference liquid kinematic viscosity m²/s
\(t_{ref}\) Reference efflux time s
\(t\) Sample efflux time s
\(\rho\) Sample density kg/m³
\(B\) Kinetic energy correction constant (mm²) -
\(C\) Viscometer constant (mm²/s²) -
\(\nu\) Sample kinematic viscosity m²/s
\(\mu\) Sample dynamic viscosity Pa·s
\(\frac{B/t^2}{C}\) Kinetic energy correction share -

Method and standards

Capillary viscometer.

Where it is used

  • Laboratory oil and fuel viscosity testing.
  • Checking viscometer constants.

Common mistakes

  • Testing at a different temperature from calibration.
  • Using a tube that gives very short efflux times.

Assumptions and limits

  • Newtonian liquid; laminar flow; same temperature for reference and sample.

References

  • Hagen–Poiseuille capillary viscometry with kinetic energy correction

Questions and answers

Why calibrate with a reference liquid?

The tube's radius enters to the fourth power; a calibration liquid fixes the constant far more accurately than measuring the capillary.

When does the kinetic energy correction matter?

For short efflux times (fast, low-viscosity flows); choose a viscometer giving more than 200 s.

What do the defaults represent?

An illustrative viscometer calibrated with water (1.0034 mm²/s at 20 °C) in 340 s, timing an oil of 880 kg/m³ at 512 s.

Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Sources

Van Wazer, Lyons, Kim and Colwell, Viscosity and Flow Measurement (Interscience, 1963), Bird, Stewart and Lightfoot, Transport Phenomena, 2nd ed. (Wiley, 2002)

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.