Worked example
Example: Illustrative oil, water-calibrated. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Reference liquid kinematic viscosity | 1.0034 cSt |
| Reference efflux time | 340.00 s |
| Sample efflux time | 512.00 s |
| Sample density | 880.00 kg/m³ |
| Kinetic energy correction constant (mm²) | 0 - |
Applying the formula below gives:
| Result | Value |
|---|---|
| Viscometer constant (mm²/s²) | 0.002951 |
| Sample kinematic viscosity | 1.5110 cSt |
| Sample dynamic viscosity | 1.3297 mPa·s |
| Kinetic energy correction share | 0 % |
Open this example in the calculator and change any input.
Formula
\(C = \frac{\nu_{ref} + B/t_{ref}}{t_{ref}}, \quad \nu = Ct - \frac Bt, \quad \mu = \rho\nu\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(\nu_{ref}\) | Reference liquid kinematic viscosity | m²/s |
| \(t_{ref}\) | Reference efflux time | s |
| \(t\) | Sample efflux time | s |
| \(\rho\) | Sample density | kg/m³ |
| \(B\) | Kinetic energy correction constant (mm²) | - |
| \(C\) | Viscometer constant (mm²/s²) | - |
| \(\nu\) | Sample kinematic viscosity | m²/s |
| \(\mu\) | Sample dynamic viscosity | Pa·s |
| \(\frac{B/t^2}{C}\) | Kinetic energy correction share | - |
Method and standards
Capillary viscometer.
Where it is used
- Laboratory oil and fuel viscosity testing.
- Checking viscometer constants.
Common mistakes
- Testing at a different temperature from calibration.
- Using a tube that gives very short efflux times.
Assumptions and limits
- Newtonian liquid; laminar flow; same temperature for reference and sample.
References
- Hagen–Poiseuille capillary viscometry with kinetic energy correction
Questions and answers
Why calibrate with a reference liquid?
The tube's radius enters to the fourth power; a calibration liquid fixes the constant far more accurately than measuring the capillary.
When does the kinetic energy correction matter?
For short efflux times (fast, low-viscosity flows); choose a viscometer giving more than 200 s.
What do the defaults represent?
An illustrative viscometer calibrated with water (1.0034 mm²/s at 20 °C) in 340 s, timing an oil of 880 kg/m³ at 512 s.
Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.