Pressure Transmitter Total Probable Error (TPE)

Total probable error of a pressure or DP transmitter from reference accuracy, ambient temperature effect and static pressure effect, combined by root sum square.

Inputs

Examples: Illustrative DP transmitter 25 of 62 kPaHigh turndown, 5 of 62 kPaGauge transmitter, no static effect

Accepted: 0.1 to 100,000 kPa
Accepted: 0.01 to 100,000 kPa
Accepted: 0 to 5 %
Accepted: 0 to 5 %
Accepted: 0 to 5 %
Accepted: 0 to 200 K
Accepted: 0 to 5 %
Accepted: 0 to 420 bar

Results

Manufacturer specifications give a reference accuracy (% span), an ambient temperature effect as (a % URL + b % span) per 28 K (50 °F), and a static pressure zero effect in % URL per 69 bar (1,000 psi). Each is converted to % of calibrated span and combined by root sum square: TPE = √(e_ref² + e_T² + e_SP²). Turning a high-URL transmitter down to a small span magnifies the URL-based terms. Span effects of static pressure that are calibrated out are excluded; long-term stability can be added as another term.

Temperature effect (% span)0.43500%
Static pressure effect (% span)1.2400%
Total probable error (% span)1.3157%
Total probable error0.065785kPa
Worst case (arithmetic sum, % span)1.7400%

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: Illustrative DP transmitter 25 of 62 kPa. The inputs, in the units selected above:

InputValue
Upper range limit62.000 kPa
Calibrated span25.000 kPa
Reference accuracy (% of span)0.065000 %
Temperature effect, URL part per 28 K0.025000 %
Temperature effect, span part per 28 K0.12500 %
Ambient change from calibration28.000 K
Static pressure zero effect per 69 bar (% URL)0.10000 %
Line static pressure69.000 bar

Applying the formula below gives:

ResultValue
Temperature effect (% span)0.18700 %
Static pressure effect (% span)0.24800 %
Total probable error (% span)0.31733 %
Total probable error0.079332 kPa
Worst case (arithmetic sum, % span)0.50000 %

Open this example in the calculator and change any input.

Formula

\(e_T = \left(a_T\frac{URL}{S} + b_T\right)\frac{\Delta T}{28}, \quad e_{SP} = z_p\frac{URL}{S}\frac{p_{st}}{69\ \text{bar}}, \quad TPE = \sqrt{e_{ref}^2 + e_T^2 + e_{SP}^2}\)

Variables

Symbol Meaning SI unit
\(URL\) Upper range limit Pa
\(S\) Calibrated span Pa
\(e_{ref}\) Reference accuracy (% of span) -
\(a_T\) Temperature effect, URL part per 28 K -
\(b_T\) Temperature effect, span part per 28 K -
\(\Delta T\) Ambient change from calibration K
\(z_p\) Static pressure zero effect per 69 bar (% URL) -
\(p_{st}\) Line static pressure Pa
\(e_T\) Temperature effect (% span) -
\(e_{SP}\) Static pressure effect (% span) -
\(TPE\) Total probable error (% span) -
\(e_{max}\) Worst case (arithmetic sum, % span) -

Method and standards

Total probable error.

Where it is used

  • Transmitter selection and range choice.
  • Flow and level measurement uncertainty.

Common mistakes

  • Ignoring the URL-based terms at high turndown.
  • Adding all terms arithmetically as the 'accuracy'.

Assumptions and limits

  • Independent errors; specifications quoted as described.

References

  • Root-sum-square combination of specification errors

Questions and answers

Why root sum square rather than adding?

The individual errors are independent and unlikely to peak together; RSS gives the probable total, the sum gives the worst case.

Why does turndown hurt accuracy?

Effects quoted in % of URL become a bigger share of a small span.

What do the defaults represent?

An illustrative 62 kPa URL DP transmitter calibrated to 25 kPa, ±0.065 % reference accuracy, 28 K ambient change, at 69 bar static pressure.

Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Sources

Lipták, Instrument Engineers' Handbook, Vol. 1: Process Measurement and Analysis, 4th ed. (CRC Press, 2003)

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.