Worked example
Example: 2 mV/V, 10 V, 1 000 kg cell at 350 kg; 1 µV. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Rated output (mV per V at rated load) | 2.0000 - |
| Excitation voltage | 10.000 V |
| Rated capacity | 1,000.0 kg |
| Applied load | 350.00 kg |
| Resolution of the indicator | 1.0000 µV |
Applying the formula below gives:
| Result | Value |
|---|---|
| Output at rated load | 20.000 mV |
| Output at the applied load | 7.0000 mV |
| Smallest load step the indicator resolves | 0.050000 kg |
Open this example in the calculator and change any input.
Formula
\(V_{FS} = S\, V_{ex}, \quad V_o = V_{FS}\, \frac{m}{m_r}\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(S\) | Rated output (mV per V at rated load) | - |
| \(V_{ex}\) | Excitation voltage | V |
| \(m_r\) | Rated capacity | kg |
| \(m\) | Applied load | kg |
| \(\delta V\) | Resolution of the indicator | V |
| \(V_{FS}\) | Output at rated load | V |
| \(V_o\) | Output at the applied load | V |
| \(\delta m\) | Smallest load step the indicator resolves | kg |
Method and standards
Definition of mV/V.
Where it is used
- Weighing system design.
- Checking indicator readings.
Common mistakes
- mV/V treated as mV.
- Several cells in parallel without averaging.
Assumptions and limits
- Linear cell.
References
- Definition of rated output (OIML R 60 terminology)
Questions and answers
A 2 mV/V, 1000 kg cell on 10 V at 350 kg?
20 mV full scale, 7 mV at 350 kg.
What does 1 µV resolution mean?
0.05 kg steps on this cell.
Why ratiometric?
The output is proportional to excitation, so the indicator divides by it.
Method and article reviewed by FireCAD engineering on 12 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.