Worked example
Example: Water-calibrated, oil 850, stainless float. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Scale reading (in the calibration fluid) | 100.00 - |
| Float density | 8,000.0 kg/m³ |
| Density of the calibration fluid | 1,000.0 kg/m³ |
| Density of the actual fluid | 850.00 kg/m³ |
Applying the formula below gives:
| Result | Value |
|---|---|
| Volume flow correction factor | 1.0962 |
| Actual volume flow | 109.62 |
| Mass flow ratio factor (k ρ₂/ρ₁) | 0.93178 |
Open this example in the calculator and change any input.
Formula
\(k = \sqrt{\frac{\rho_1 (\rho_f - \rho_2)}{\rho_2 (\rho_f - \rho_1)}}, \quad Q_v = k\, Q_s\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(Q_s\) | Scale reading (in the calibration fluid) | - |
| \(\rho_f\) | Float density | kg/m³ |
| \(\rho_1\) | Density of the calibration fluid | kg/m³ |
| \(\rho_2\) | Density of the actual fluid | kg/m³ |
| \(k\) | Volume flow correction factor | - |
| \(Q_v\) | Actual volume flow | - |
| \(Q_m\) | Mass flow ratio factor (k ρ₂/ρ₁) | - |
Method and standards
Drag on the float balances its submerged weight.
Where it is used
- Purge and sample flows.
- Reusing a meter on another fluid.
Common mistakes
- Gas density at standard instead of meter conditions.
- Ignoring viscosity for oils.
Assumptions and limits
- Same viscosity regime.
References
- Force balance on the float of a variable-area meter (ISO 11605 principle)
Questions and answers
What does a water-calibrated rotameter read for an 850 kg/m³ oil?
Volume flow = scale × 1.0962 with a stainless float.
Why does float density matter?
The float hangs where drag balances its weight less buoyancy.
And for gases?
The float density is far above the gas density, so k ≈ √(ρ₁/ρ₂).
Method and article reviewed by FireCAD engineering on 12 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.