Worked example
Example: Water 10 kg/s in DN100 (102.3 mm) at 250 mbar, flange taps. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Pipe inside diameter | 102.30 mm |
| Maximum mass flow | 36,000 kg/h |
| Differential pressure wanted at maximum flow | 250.00 mbar |
| Pressure tappings | Flange |
| Upstream pressure (absolute) | 5.0000 bar |
| Density at the upstream tapping | 998.20 kg/m³ |
| Dynamic viscosity | 1.0020 cP |
| Isentropic exponent (0 = liquid) | 0 - |
Applying the formula below gives:
| Result | Value |
|---|---|
| Orifice bore | 53.456 mm |
| Diameter ratio d/D | 0.52254 |
| Discharge coefficient | 0.60673 |
| Expansibility factor | 1.0000 |
| Permanent pressure loss | 177.45 mbar |
Open this example in the calculator and change any input.
Formula
\(q_m = \frac{C}{\sqrt{1 - \beta^4}}\,\varepsilon\,\frac{\pi}{4}d^2\sqrt{2\,\Delta p\,\rho_1} \qquad \beta = \frac{d}{D} \text{ found by bisection}\)
\(\varepsilon = 1 - (0.351 + 0.256\beta^4 + 0.93\beta^8)\left[1 - \left(\frac{p_2}{p_1}\right)^{1/\kappa}\right], \quad \Delta\varpi = \frac{\sqrt{1 - \beta^4(1 - C^2)} - C\beta^2}{\sqrt{1 - \beta^4(1 - C^2)} + C\beta^2}\,\Delta p\)
C is the Reader-Harris/Gallagher equation of ISO 5167-2 (shown in full on the orifice flow calculator), a function of β, the Reynolds number and the tapping positions.
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(D\) | pipe inside diameter | m |
| \(q_m\) | maximum mass flow | kg/s |
| \(\Delta p\) | differential pressure at max flow | Pa |
| \(\text{taps}\) | tapping arrangement | – |
| \(p_1\) | upstream pressure | Pa |
| \(\rho_1\) | upstream density | kg/m³ |
| \(\mu\) | viscosity | Pa·s |
| \(\kappa\) | isentropic exponent (0 liquid) | – |
| \(d\) | bore | m |
| \(\beta\) | diameter ratio | – |
| \(C\) | discharge coefficient | – |
| \(\varepsilon\) | expansibility | – |
| \(\Delta\varpi\) | permanent pressure loss | Pa |
Method and standards
ISO 5167-2 orifice equation with the Reader-Harris/Gallagher discharge coefficient and the expansibility equation, solved for β by bisection.
Where it is used
- Feedwater, steam and fuel flow elements.
- Replacing a damaged plate.
- Re-ranging a DP transmitter.
Common mistakes
- Choosing β above 0.75.
- Using nominal pipe size.
- Forgetting straight-length requirements.
Assumptions and limits
- The bore is searched between β 0.05 and 0.85, wider than the ISO 5167-2 range of 0.10 to 0.75; a result outside that range is given with a warning and is outside the standard's uncertainty statement.
- Within ISO 5167-2 limits; straight lengths as the standard requires.
References
- ISO 5167-1
- ISO 5167-2
Questions and answers
What bore gives 250 mbar at 10 kg/s of water in DN100?
53.5 mm (β 0.52) with flange taps; C 0.607 and about 178 mbar permanent loss.
Which differential pressure should I choose?
Commonly 250–500 mbar for liquids and steam; lower for low-pressure gases. Keep β within 0.2–0.7 for best accuracy.
What about steam?
Enter density at the upstream condition and the isentropic exponent (1.3 superheated); expansibility is included.