Worked example
Example: 50 120 pulses for 5 m³ at K 10 000/m³. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Pulses counted in the run | 50,120 - |
| True volume (prover or tank) | 5.0000 m³ |
| K-factor programmed in the meter, in pulses per m³ (per litre × 1000; per US gal × 264.17) | 10,000 - |
Applying the formula below gives:
| Result | Value |
|---|---|
| Measured K-factor (pulses per m³) | 10,024 |
| Volume indicated by the meter | 5.0120 m³ |
| Meter factor (true / indicated) | 0.99761 |
| Meter error (indicated − true) / true | 0.24000 % |
Open this example in the calculator and change any input.
Formula
\(K = \frac{N}{V_t}, \quad V_i = \frac{N}{K_0}, \quad MF = \frac{V_t}{V_i}\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(N\) | Pulses counted in the run | - |
| \(V_t\) | True volume (prover or tank) | m³ |
| \(K_0\) | K-factor programmed in the meter (pulses per m³) | - |
| \(K\) | Measured K-factor (pulses per m³) | - |
| \(V_i\) | Volume indicated by the meter | m³ |
| \(MF\) | Meter factor (true / indicated) | - |
| \(\varepsilon\) | Meter error (indicated − true) / true | - |
Method and standards
Definitions.
Where it is used
- Meter calibration and custody transfer.
- Tank-strapping checks.
Common mistakes
- Mixing per litre and per m³.
- Too few pulses.
Assumptions and limits
- Steady flow during the run.
- No temperature and pressure corrections (CTL, CPL) of the prover and meter volumes as in API MPMS proving; for liquids away from reference conditions apply them first.
- K-factors in pulses per m³: multiply a per-litre figure by 1 000 or a per-US-gallon figure by 264.17.
References
- Meter factor MF = true volume / indicated volume (API MPMS ch. 4 terminology)
Questions and answers
50 120 pulses for 5 m³ on a meter set to 10 000/m³?
K = 10 024/m³; the meter shows 5.012 m³, reading 0.24 % high; meter factor 0.9976.
Change K or apply a meter factor?
Either; custody transfer usually keeps K and applies the factor.
How many pulses per run?
At least 10 000 so one pulse is 0.01 %.