Standard Addition: Sample Concentration from Spiked Signals

Unknown concentration by the standard addition method from three to five signals at added concentrations, by linear regression extrapolated to zero signal.

Process steam and utilitiesRefinery and petrochemicalChemical and fertiliser

Inputs

Examples: Illustrative four additions, two-fold dilutionThree additions, no dilutionFive additions, ten-fold dilution

Accepted: 3 to 5
Accepted: 1 to 10,000
Accepted: 0 to 1,000,000 mg/L
Accepted: 0 to 1E+12
Accepted: 0 to 1,000,000 mg/L
Accepted: 0 to 1E+12
Accepted: 0 to 1,000,000 mg/L
Accepted: 0 to 1E+12
Accepted: 0 to 1,000,000 mg/L
Accepted: 0 to 1E+12
Accepted: 0 to 1,000,000 mg/L
Accepted: 0 to 1E+12

Results

Signal y = m(C_m + x): the x-intercept −b/m gives the concentration in the measured solution, C_m = b/m; multiply by the dilution factor for the original sample. Standard addition corrects matrix effects that change sensitivity, but not additive interferences or a non-zero blank.

Slope (signal per concentration)102.70
Intercept0.21570
Concentration in measured solution2.1003mg/L
Concentration in original sample4.2006mg/L
Coefficient of determination0.99977

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: Illustrative four additions, two-fold dilution. The inputs, in the units selected above:

InputValue
Number of points4.0000 -
Dilution factor (final volume / sample volume)2.0000 -
Added 1 (usually 0)0 mg/L
Signal 10.21500 -
Added 21.0000 mg/L
Signal 20.31800 -
Added 32.0000 mg/L
Signal 30.42400 -
Added 43.0000 mg/L
Signal 40.52200 -
Added 54.0000 mg/L
Signal 50.63000 -

Applying the formula below gives:

ResultValue
Slope (signal per concentration)102.70
Intercept0.21570
Concentration in measured solution2.1003 mg/L
Concentration in original sample4.2006 mg/L
Coefficient of determination0.99977

Open this example in the calculator and change any input.

Formula

\(y = m(C_m + x) = b + mx, \quad C_m = \frac{b}{m}, \quad C_0 = f_dC_m\)

Variables

Symbol Meaning SI unit
\(n\) Number of points -
\(f_d\) Dilution factor (final volume / sample volume) -
\(x_1\) Added 1 (usually 0) kg/m³
\(y_1\) Signal 1 -
\(x_2\) Added 2 kg/m³
\(y_2\) Signal 2 -
\(x_3\) Added 3 kg/m³
\(y_3\) Signal 3 -
\(x_4\) Added 4 kg/m³
\(y_4\) Signal 4 -
\(x_5\) Added 5 kg/m³
\(y_5\) Signal 5 -
\(m\) Slope (signal per concentration) -
\(b\) Intercept -
\(C_m\) Concentration in measured solution kg/m³
\(C_0\) Concentration in original sample kg/m³
\(R^2\) Coefficient of determination -

Method and standards

Standard addition.

Where it is used

  • Trace metals in complex matrices.
  • Ion-selective electrode measurements.

Common mistakes

  • Forgetting to correct for dilution by the spikes.
  • Spiking too little to change the signal clearly.

Assumptions and limits

  • Linear response; no blank signal.

References

  • Standard addition method

Questions and answers

When is standard addition used?

When the sample matrix changes instrument sensitivity, as in AAS of brines, ICP of digests or electrodes in effluent.

Why must the response be linear?

The result is an extrapolation beyond the data; curvature gives large errors.

What do the defaults represent?

Four illustrative signals for 0, 1, 2 and 3 mg/L added to a sample diluted two-fold.

Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Sources

Harris, Quantitative Chemical Analysis, 9th ed. (Freeman, 2015)

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.