Worked example
Example: 300 and 325 s on 30 m. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Retention time, peak 1 | 300.00 s |
| Base width, peak 1 | 12.000 s |
| Retention time, peak 2 | 325.00 s |
| Base width, peak 2 | 13.000 s |
| Column length | 30.000 m |
Applying the formula below gives:
| Result | Value |
|---|---|
| Theoretical plates (peak 2) | 10,000 |
| Plate height (HETP) | 3.0000 mm |
| Resolution | 2.0000 |
Open this example in the calculator and change any input.
Formula
\(N = 16\left(\frac{t_R}{w}\right)^2, \quad H = \frac LN, \quad R_s = \frac{2(t_{R2} - t_{R1})}{w_1 + w_2}\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(t_{R1}\) | Retention time, peak 1 | s |
| \(w_1\) | Base width, peak 1 | s |
| \(t_{R2}\) | Retention time, peak 2 | s |
| \(w_2\) | Base width, peak 2 | s |
| \(L\) | Column length | m |
| \(N\) | Theoretical plates (peak 2) | - |
| \(H\) | Plate height (HETP) | m |
| \(R_s\) | Resolution | - |
Method and standards
Chromatography definitions.
Where it is used
- Process GC method checks.
- Column performance tracking.
Common mistakes
- Using half-height width with the base-width formula (use 5.54 instead of 16).
- Mixing time units.
Assumptions and limits
- Gaussian peaks.
References
- N = 16(t_R/w)²
- R_s = 2(t_R2 − t_R1)/(w₁ + w₂) (analytical chemistry definitions)
Questions and answers
What resolution is enough?
1.5 gives baseline separation of similar-sized peaks; process GC methods often target 1.5–2.
How can resolution be improved?
Longer columns (R_s grows with √N), better selectivity (a different phase or temperature) or more retention.
What do the defaults represent?
Two peaks at 300 and 325 s on a 30 m column.