pH Electrode Nernst Slope and Temperature Compensation

Theoretical slope of a pH electrode at a temperature from the Nernst equation, the electrode voltage at a pH, and the pH from a measured voltage.

Thermal power plantRefinery and petrochemicalChemical and fertiliser

Inputs

Examples: 25 °C, pH 4, −100 mV reading80 °C, pH 9.5, +150 mV, 97 % slope5 °C, pH 7, 0 mV

Results

Automatic temperature compensation corrects the electrode slope only; the sample's own pH also changes with temperature.

Slope at this temperature (mV per pH)59.159mV
Voltage at the pH entered177.48mV
pH at the voltage entered8.6903

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: 25 °C, pH 4, −100 mV reading. The inputs, in the units selected above:

InputValue
Temperature of the solution25.000 °C
pH (to find the voltage)4.0000 -
Measured electrode voltage (to find the pH)-100.00 mV
Isopotential (zero-voltage) pH7.0000 -
Electrode slope efficiency100.00 %

Applying the formula below gives:

ResultValue
Slope at this temperature (mV per pH)59.159 mV
Voltage at the pH entered177.48 mV
pH at the voltage entered8.6903

Open this example in the calculator and change any input.

Formula

\(S = \eta_s\, \frac{\ln 10\, R\, T}{F}, \quad E = -S\,(pH - pH_0)\)

Variables

Symbol Meaning SI unit
\(T\) Temperature of the solution K
\(pH\) pH (to find the voltage) -
\(E\) Measured electrode voltage (to find the pH) V
\(pH_0\) Isopotential (zero-voltage) pH -
\(\eta_s\) Electrode slope efficiency -
\(S\) Slope at this temperature (mV per pH) V
\(E_{pH}\) Voltage at the pH entered V
\(pH_E\) pH at the voltage entered -

Method and standards

Nernst equation for a glass electrode.

Where it is used

  • Boiler water and effluent pH loops.
  • Calibration checks.

Common mistakes

  • Compensating the sample's chemistry with ATC.
  • Old electrodes with low slope.

Assumptions and limits

  • Ideal glass electrode.

References

  • Nernst equation
  • slope = ln(10) R T / F (CODATA 2018 R and F)

Questions and answers

What is the slope at 25 °C?

59.16 mV per pH.

And at 80 °C?

70.07 mV per pH: why ATC is needed.

What slope efficiency is good?

95 to 102 % after a two-buffer calibration; lower means an ageing electrode.

Method and article reviewed by FireCAD engineering on 12 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.