Worked example
Example: Illustrative NADH, 35 % T. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Transmittance | 35.000 % |
| Molar absorptivity (L/(mol·cm)) | 6,220.0 - |
| Path length | 1.0000 cm |
| Molar mass | 663.40 kg/kmol |
Applying the formula below gives:
| Result | Value |
|---|---|
| Absorbance | 0.45593 |
| Molar concentration | 0.073301 mmol/L |
| Mass concentration | 48.628 mg/L |
Open this example in the calculator and change any input.
Formula
\(A = -\log_{10}T = \varepsilon lc, \quad c = \frac{A}{\varepsilon l}\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(T\) | Transmittance | - |
| \(\varepsilon\) | Molar absorptivity (L/(mol·cm)) | - |
| \(l\) | Path length | m |
| \(M\) | Molar mass | kg/kmol |
| \(A\) | Absorbance | - |
| \(c\) | Molar concentration | mol/m³ |
| \(c_m\) | Mass concentration | kg/m³ |
Method and standards
Absorbance.
Where it is used
- Process photometers and colour analysers.
- Laboratory UV-visible assays.
Common mistakes
- Using percent transmittance in the logarithm.
- Ignoring the blank.
Assumptions and limits
- Monochromatic light; dilute, non-scattering sample.
References
- Beer–Lambert–Bouguer law
Questions and answers
Why does the calibration curve bend at high absorbance?
Stray light, polychromatic light and molecular interactions break the linear law.
What is a typical path length?
1 cm cuvettes in the laboratory; process photometers use from millimetres to tens of centimetres.
What do the defaults represent?
An illustrative NADH-type measurement: 35 % transmittance in a 1 cm cell with ε = 6,220 L/(mol·cm) and M = 663.4 g/mol.
Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.
Sources
Skoog, Holler and Crouch, Principles of Instrumental Analysis, 7th ed. (Cengage, 2017)