Worked example
Example: Illustrative sodium electrode with potassium. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Measured potential | 118.00 mV |
| Electrode intercept (from calibration) | 295.00 mV |
| Charge of the primary ion | 1.0000 - |
| Charge of the interfering ion | 1.0000 - |
| Selectivity coefficient | 0.010000 - |
| Interfering ion activity (mol/L) | 0.050000 - |
| Temperature | 25.000 °C |
Applying the formula below gives:
| Result | Value |
|---|---|
| Nernst slope per decade | 59.159 mV |
| Apparent activity (no correction, mol/L) | 0.001019 |
| Primary ion activity (mol/L) | 5.1878e-04 |
| Error if interference ignored | 96.380 % |
Open this example in the calculator and change any input.
Formula
\(E = E^0 + \frac{2.303RT}{z_iF}\log\left(a_i + K_{ij}a_j^{z_i/z_j}\right)\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(E\) | Measured potential | V |
| \(E^0\) | Electrode intercept (from calibration) | V |
| \(z_i\) | Charge of the primary ion | - |
| \(z_j\) | Charge of the interfering ion | - |
| \(K_{ij}\) | Selectivity coefficient | - |
| \(a_j\) | Interfering ion activity (mol/L) | - |
| \(T\) | Temperature | K |
| \(S\) | Nernst slope per decade | V |
| \(a_{app}\) | Apparent activity (no correction, mol/L) | - |
| \(a_i\) | Primary ion activity (mol/L) | - |
| \(a_{app}/a_i - 1\) | Error if interference ignored | - |
Method and standards
ISE interference.
Where it is used
- Water and process analysers with ion-selective electrodes.
- Choosing electrodes for a matrix.
Common mistakes
- Ignoring interference in mixed matrices.
- Confusing activity and concentration.
Assumptions and limits
- Nernstian electrode; fixed ionic strength; known selectivity coefficient.
References
- Nikolsky–Eisenman equation
Questions and answers
What is the selectivity coefficient?
How strongly the electrode responds to the interfering ion relative to the primary ion; smaller is better.
Why use an ionic strength buffer?
Electrodes respond to activity; a buffer fixes the activity coefficient so readings track concentration.
What do the defaults represent?
An illustrative sodium electrode with E⁰ 295.0 mV reading 118.0 mV, with 0.05 mol/L potassium at K = 0.01, 25 °C.
Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.
Sources
Buck and Lindner, Recommendations for Nomenclature of Ion-Selective Electrodes, Pure Appl. Chem. 66 (1994), Bard and Faulkner, Electrochemical Methods, 2nd ed. (Wiley, 2001)