Worked example
Example: Illustrative oil sample. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Water in the standard | 30.000 mg |
| Titrant used on the standard | 6.0200 mL |
| Titrant used on the sample | 3.4500 mL |
| Blank (drift) volume | 0.050000 mL |
| Sample mass | 2.0000 g |
| Coulometric charge (0 to skip) | 3.3333e-04 Ah |
Applying the formula below gives:
| Result | Value |
|---|---|
| Titer (mg water per mL) | 0.004983 |
| Water found | 16.944 mg |
| Water content | 0.84718 % |
| Water content (ppm by mass) | 8,471.8 |
| Water from the coulometric charge | 0.11203 mg |
Open this example in the calculator and change any input.
Formula
\(F = \frac{m_{std}}{V_{std}}, \quad w = \frac{(V - V_b)F}{m}, \quad m_{w,c} = \frac{QM_w}{2F_c}\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(m_{std}\) | Water in the standard | kg |
| \(V_{std}\) | Titrant used on the standard | m³ |
| \(V\) | Titrant used on the sample | m³ |
| \(V_b\) | Blank (drift) volume | m³ |
| \(m\) | Sample mass | kg |
| \(Q\) | Coulometric charge (0 to skip) | C |
| \(F\) | Titer (mg water per mL) | - |
| \(m_w\) | Water found | kg |
| \(w\) | Water content | - |
| \(m_{w,c}\) | Water from the coulometric charge | kg |
Method and standards
Karl Fischer.
Where it is used
- Moisture in oils, solvents and transformer oil.
- Pharmaceutical and food water content.
Common mistakes
- Forgetting the blank.
- Mixing titer units (mg/mL).
Assumptions and limits
- Stoichiometric KF reaction; no side reactions.
References
- Karl Fischer stoichiometry (1 H₂O : 1 I₂ : 2 e⁻)
Questions and answers
Why determine the titer often?
Volumetric reagents lose strength with time and moisture pick-up.
When is coulometric better?
For small amounts of water (ppm levels) because it needs no titer and measures micrograms.
What do the defaults represent?
An illustrative titration: 30 mg water standard uses 6.02 mL; a 2 g oil sample uses 3.45 mL with 0.05 mL blank; 1.2 C coulometric check.
Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.
Sources
Scholz, Karl Fischer Titration (Springer, 1984), Lipták (ed.), Instrument Engineers' Handbook, Vol. 1: Process Measurement and Analysis, 4th ed. (CRC Press, 2003)