FireCADCalc

pH Meter Two-Point Calibration: Slope, Offset and Sample pH

Electrode slope, percent of the Nernst slope and zero-point offset from two buffer readings, and the pH of a sample from its millivolt reading.

Inputs

Examples: Illustrative pH 7 and 4 buffers at 25 CAlkaline range, pH 7 and 10 at 25 CHot sample line at 50 C

Accepted: 0 to 14
Accepted: -1,000 to 1,000 mV
Accepted: 0 to 14
Accepted: -1,000 to 1,000 mV
Accepted: -10 to 150 °C
Accepted: -1,000 to 1,000 mV

Results

The glass electrode follows E = E₀ − s(pH − 7). Two buffers fix the slope s (mV per pH, positive as entered) and the offset at pH 7. The theoretical slope is ln10·RT/F (59.16 mV/pH at 25 °C). A healthy electrode shows 95–102 % slope and an offset within about ±30 mV; lower slopes mean an ageing or dirty membrane. Measure the sample at the calibration temperature, or apply temperature compensation.

Electrode slope56.689mV
Theoretical Nernst slope59.159mV
Slope efficiency95.824%
Offset (reading at pH 7)-3.0000mV
Sample pH6.0651

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: Illustrative pH 7 and 4 buffers at 25 C. The inputs, in the units selected above:

InputValue
Buffer 1 pH7.0000 -
Reading in buffer 1-3.0000 mV
Buffer 2 pH4.0100 -
Reading in buffer 2166.50 mV
Calibration temperature25.000 °C
Sample reading50.000 mV

Applying the formula below gives:

ResultValue
Electrode slope56.689 mV
Theoretical Nernst slope59.159 mV
Slope efficiency95.824 %
Offset (reading at pH 7)-3.0000 mV
Sample pH6.0651

Open this example in the calculator and change any input.

Formula

\(s = \frac{E_2 - E_1}{pH_1 - pH_2}, \quad s_N = \frac{\ln10\,RT}{F}, \quad pH_x = 7 - \frac{E_x - E_{pH7}}{s}\)

Variables

Symbol Meaning SI unit
\(pH_1\) Buffer 1 pH -
\(E_1\) Reading in buffer 1 V
\(pH_2\) Buffer 2 pH -
\(E_2\) Reading in buffer 2 V
\(T\) Calibration temperature K
\(E_x\) Sample reading V
\(s\) Electrode slope V
\(s_N\) Theoretical Nernst slope V
\(s/s_N\) Slope efficiency -
\(E_{pH7}\) Offset (reading at pH 7) V
\(pH_x\) Sample pH -

Method and standards

pH calibration.

Where it is used

  • Routine analyser and laboratory meter calibration.
  • Electrode health checks.

Common mistakes

  • Calibrating at one temperature and measuring at another.
  • Using expired or contaminated buffers.

Assumptions and limits

  • Linear electrode response between buffers; buffers at stated pH for the temperature.

References

  • Two-point calibration of a glass electrode

Questions and answers

What slope is acceptable?

Typically 95–102 % of the Nernst value; below about 90 % clean or replace the electrode.

Why bracket the sample with buffers?

The electrode is closest to linear between the buffers; extrapolating far outside them adds error.

What do the defaults represent?

An illustrative calibration at 25 °C with pH 7.00 and 4.01 buffers reading −3 and +166.5 mV, and a sample at +50 mV.

Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Sources

Lipták, Instrument Engineers' Handbook, Vol. 1: Process Measurement and Analysis, 4th ed. (CRC Press, 2003)

Related calculators