Worked example
Example: Illustrative Fe3+/Fe2+ at pH 2. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Standard potential vs SHE | 771.00 mV |
| Electrons transferred | 1.0000 - |
| Protons consumed in the reduction | 0 - |
| Oxidised species activity | 0.001000 - |
| Reduced species activity | 1.0000e-04 - |
| pH | 2.0000 - |
| Temperature | 25.000 °C |
| Reference electrode potential vs SHE | 197.00 mV |
Applying the formula below gives:
| Result | Value |
|---|---|
| Slope per decade | 59.159 mV |
| Potential vs SHE | 830.16 mV |
| Reading vs the reference electrode | 633.16 mV |
Open this example in the calculator and change any input.
Formula
\(E_h = E^0 + \frac{2.303RT}{nF}\log\frac{a_{ox}}{a_{red}} - \frac{2.303RT\,m}{nF}pH, \quad E_{meas} = E_h - E_{ref}\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(E^0\) | Standard potential vs SHE | V |
| \(n\) | Electrons transferred | - |
| \(m\) | Protons consumed in the reduction | - |
| \(a_{ox}\) | Oxidised species activity | - |
| \(a_{red}\) | Reduced species activity | - |
| \(pH\) | pH | - |
| \(T\) | Temperature | K |
| \(E_{ref}\) | Reference electrode potential vs SHE | V |
| \(S\) | Slope per decade | V |
| \(E_h\) | Potential vs SHE | V |
| \(E_{meas}\) | Reading vs the reference electrode | V |
Method and standards
ORP.
Where it is used
- Cyanide, chrome and chlorination control.
- Water quality monitoring.
Common mistakes
- Comparing Ag/AgCl readings with SHE values.
- Treating ORP as a concentration measurement.
Assumptions and limits
- Equilibrium; activities known; reversible couple.
References
- Nernst equation for a redox couple
Questions and answers
Why does my ORP reading differ from the table value?
Meters read against Ag/AgCl, about 200 mV below the hydrogen electrode, and real solutions are rarely at equilibrium.
Does pH affect ORP?
Yes, for reactions involving hydrogen ions: about 59 mV × m/n per pH unit at 25 °C.
What do the defaults represent?
The iron(III)/iron(II) couple (E⁰ 0.771 V) with a 10:1 ratio at pH 2 and 25 °C, read against Ag/AgCl (0.197 V).
Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.