Worked example
Example: Illustrative four additions, two-fold dilution. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Number of points | 4.0000 - |
| Dilution factor (final volume / sample volume) | 2.0000 - |
| Added 1 (usually 0) | 0 mg/L |
| Signal 1 | 0.21500 - |
| Added 2 | 1.0000 mg/L |
| Signal 2 | 0.31800 - |
| Added 3 | 2.0000 mg/L |
| Signal 3 | 0.42400 - |
| Added 4 | 3.0000 mg/L |
| Signal 4 | 0.52200 - |
| Added 5 | 4.0000 mg/L |
| Signal 5 | 0.63000 - |
Applying the formula below gives:
| Result | Value |
|---|---|
| Slope (signal per concentration) | 102.70 |
| Intercept | 0.21570 |
| Concentration in measured solution | 2.1003 mg/L |
| Concentration in original sample | 4.2006 mg/L |
| Coefficient of determination | 0.99977 |
Open this example in the calculator and change any input.
Formula
\(y = m(C_m + x) = b + mx, \quad C_m = \frac{b}{m}, \quad C_0 = f_dC_m\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(n\) | Number of points | - |
| \(f_d\) | Dilution factor (final volume / sample volume) | - |
| \(x_1\) | Added 1 (usually 0) | kg/m³ |
| \(y_1\) | Signal 1 | - |
| \(x_2\) | Added 2 | kg/m³ |
| \(y_2\) | Signal 2 | - |
| \(x_3\) | Added 3 | kg/m³ |
| \(y_3\) | Signal 3 | - |
| \(x_4\) | Added 4 | kg/m³ |
| \(y_4\) | Signal 4 | - |
| \(x_5\) | Added 5 | kg/m³ |
| \(y_5\) | Signal 5 | - |
| \(m\) | Slope (signal per concentration) | - |
| \(b\) | Intercept | - |
| \(C_m\) | Concentration in measured solution | kg/m³ |
| \(C_0\) | Concentration in original sample | kg/m³ |
| \(R^2\) | Coefficient of determination | - |
Method and standards
Standard addition.
Where it is used
- Trace metals in complex matrices.
- Ion-selective electrode measurements.
Common mistakes
- Forgetting to correct for dilution by the spikes.
- Spiking too little to change the signal clearly.
Assumptions and limits
- Linear response; no blank signal.
References
- Standard addition method
Questions and answers
When is standard addition used?
When the sample matrix changes instrument sensitivity, as in AAS of brines, ICP of digests or electrodes in effluent.
Why must the response be linear?
The result is an extrapolation beyond the data; curvature gives large errors.
What do the defaults represent?
Four illustrative signals for 0, 1, 2 and 3 mg/L added to a sample diluted two-fold.