FireCADCalc

Electrolyte Conductivity by Kohlrausch's Law and Onsager

Limiting molar conductivity from ionic conductivities, the Kohlrausch slope by the Onsager limiting law, and the conductivity of a dilute 1:1 electrolyte solution.

Inputs

Examples: Illustrative NaCl 1 mmol/LHCl trace 0.1 mmol/LKCl 0.5 mmol/L

Accepted: 1 to 400
Accepted: 1 to 400
Accepted: 1E-6 to 100 mmol/L

Results

Kohlrausch: ions migrate independently at infinite dilution, Λ⁰ = λ⁰₊ + λ⁰₋, and Λ_m = Λ⁰ − 𝒦√c for strong electrolytes. Onsager's limiting law for a 1:1 electrolyte in water at 25 °C gives 𝒦 = A + BΛ⁰ with A = 60.20 S·cm²/mol·(mol/L)^−½ and B = 0.229 (mol/L)^−½. κ = Λ_mc. Reliable below about 1 mmol/L; above about 10 mmol/L it overestimates the fall. 25 °C values; conductivity rises about 2 %/K.

Limiting molar conductivity (S·cm²/mol)426.00
Kohlrausch slope, Onsager (S·cm²/mol per (mol/L)^½)157.75
Molar conductivity at c (S·cm²/mol)424.42
Solution conductivity42.442µS/cm
Conductivity ignoring interionic effects42.600µS/cm

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: Illustrative NaCl 1 mmol/L. The inputs, in the units selected above:

InputValue
Cation limiting ionic conductivity (S·cm²/mol)50.100 -
Anion limiting ionic conductivity (S·cm²/mol)76.350 -
Concentration1.0000 mmol/L

Applying the formula below gives:

ResultValue
Limiting molar conductivity (S·cm²/mol)126.45
Kohlrausch slope, Onsager (S·cm²/mol per (mol/L)^½)89.157
Molar conductivity at c (S·cm²/mol)123.63
Solution conductivity123.63 µS/cm
Conductivity ignoring interionic effects126.45 µS/cm

Open this example in the calculator and change any input.

Formula

\(\Lambda_m^0 = \lambda_+^0 + \lambda_-^0, \quad \Lambda_m = \Lambda_m^0 - (A + B\Lambda_m^0)\sqrt c, \quad \kappa = \Lambda_mc\)

Variables

Symbol Meaning SI unit
\(\lambda_+^0\) Cation limiting ionic conductivity (S·cm²/mol) -
\(\lambda_-^0\) Anion limiting ionic conductivity (S·cm²/mol) -
\(c\) Concentration mol/m³
\(\Lambda_m^0\) Limiting molar conductivity (S·cm²/mol) -
\(\mathcal K\) Kohlrausch slope, Onsager (S·cm²/mol per (mol/L)^½) -
\(\Lambda_m\) Molar conductivity at c (S·cm²/mol) -
\(\kappa\) Solution conductivity S/m
\(\kappa^0\) Conductivity ignoring interionic effects S/m

Method and standards

Kohlrausch.

Where it is used

  • Relating conductivity to trace salt or acid levels.
  • Checking conductivity standards.

Common mistakes

  • Using mmol/L where mol/L is needed under the root.
  • Applying it to concentrated or weak electrolytes.

Assumptions and limits

  • 1:1 strong electrolyte; water at 25 °C; dilute solutions.

References

  • Kohlrausch's law and the Debye–Hückel–Onsager limiting law

Questions and answers

Why does molar conductivity fall with concentration?

Each ion's oppositely charged atmosphere drags on it (relaxation and electrophoretic effects), growing with √c.

Can I use this for boiler or condensate water?

Yes for traces of a single salt or acid, which is how conductivity is related to contaminant level after cation exchange.

What do the defaults represent?

Sodium chloride at 1 mmol/L and 25 °C (λ⁰ 50.10 for Na⁺ and 76.35 for Cl⁻).

Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Sources

Atkins and de Paula, Physical Chemistry, 9th ed. (Oxford, 2010), Lipták, Instrument Engineers' Handbook, Vol. 1: Process Measurement and Analysis, 4th ed. (CRC Press, 2003)

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.