Worked example
Example: Illustrative NaCl 1 mmol/L. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Cation limiting ionic conductivity (S·cm²/mol) | 50.100 - |
| Anion limiting ionic conductivity (S·cm²/mol) | 76.350 - |
| Concentration | 1.0000 mmol/L |
Applying the formula below gives:
| Result | Value |
|---|---|
| Limiting molar conductivity (S·cm²/mol) | 126.45 |
| Kohlrausch slope, Onsager (S·cm²/mol per (mol/L)^½) | 89.157 |
| Molar conductivity at c (S·cm²/mol) | 123.63 |
| Solution conductivity | 123.63 µS/cm |
| Conductivity ignoring interionic effects | 126.45 µS/cm |
Open this example in the calculator and change any input.
Formula
\(\Lambda_m^0 = \lambda_+^0 + \lambda_-^0, \quad \Lambda_m = \Lambda_m^0 - (A + B\Lambda_m^0)\sqrt c, \quad \kappa = \Lambda_mc\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(\lambda_+^0\) | Cation limiting ionic conductivity (S·cm²/mol) | - |
| \(\lambda_-^0\) | Anion limiting ionic conductivity (S·cm²/mol) | - |
| \(c\) | Concentration | mol/m³ |
| \(\Lambda_m^0\) | Limiting molar conductivity (S·cm²/mol) | - |
| \(\mathcal K\) | Kohlrausch slope, Onsager (S·cm²/mol per (mol/L)^½) | - |
| \(\Lambda_m\) | Molar conductivity at c (S·cm²/mol) | - |
| \(\kappa\) | Solution conductivity | S/m |
| \(\kappa^0\) | Conductivity ignoring interionic effects | S/m |
Method and standards
Kohlrausch.
Where it is used
- Relating conductivity to trace salt or acid levels.
- Checking conductivity standards.
Common mistakes
- Using mmol/L where mol/L is needed under the root.
- Applying it to concentrated or weak electrolytes.
Assumptions and limits
- 1:1 strong electrolyte; water at 25 °C; dilute solutions.
References
- Kohlrausch's law and the Debye–Hückel–Onsager limiting law
Questions and answers
Why does molar conductivity fall with concentration?
Each ion's oppositely charged atmosphere drags on it (relaxation and electrophoretic effects), growing with √c.
Can I use this for boiler or condensate water?
Yes for traces of a single salt or acid, which is how conductivity is related to contaminant level after cation exchange.
What do the defaults represent?
Sodium chloride at 1 mmol/L and 25 °C (λ⁰ 50.10 for Na⁺ and 76.35 for Cl⁻).
Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.
Sources
Atkins and de Paula, Physical Chemistry, 9th ed. (Oxford, 2010), Lipták, Instrument Engineers' Handbook, Vol. 1: Process Measurement and Analysis, 4th ed. (CRC Press, 2003)