FireCADCalc

Karl Fischer Water Content: Volumetric and Coulometric

Titer of a volumetric Karl Fischer reagent from a water standard, water content of a sample in percent and ppm, and the water equivalent of a coulometric charge.

Inputs

Examples: Illustrative oil sampleSolvent, low waterPowder with 4 % water

Accepted: 0.1 to 1,000 mg
Accepted: 0.01 to 1,000 mL
Accepted: 0 to 1,000 mL
Accepted: 0 to 100 mL
Accepted: 0.001 to 100 g
Accepted: 0 to 0.027778 Ah

Results

Volumetric: the titer F (mg water per mL) comes from a weighed water standard; water = (V − V_blank)·F and content = water/sample mass. Coulometric: iodine is generated electrically and one water molecule consumes one I₂, i.e. two electrons, so water = Q·M_w/(2F) (about 0.0933 mg per coulomb). Coulometry suits 10 µg–10 mg of water; volumetric suits larger amounts. Side reactions (ketones, aldehydes, oxidisers) need special reagents.

Titer (mg water per mL)0.004950
Water found40.297mg
Water content4.0297%
Water content (ppm by mass)40,297
Water from the coulometric charge0mg

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: Illustrative oil sample. The inputs, in the units selected above:

InputValue
Water in the standard30.000 mg
Titrant used on the standard6.0200 mL
Titrant used on the sample3.4500 mL
Blank (drift) volume0.050000 mL
Sample mass2.0000 g
Coulometric charge (0 to skip)3.3333e-04 Ah

Applying the formula below gives:

ResultValue
Titer (mg water per mL)0.004983
Water found16.944 mg
Water content0.84718 %
Water content (ppm by mass)8,471.8
Water from the coulometric charge0.11203 mg

Open this example in the calculator and change any input.

Formula

\(F = \frac{m_{std}}{V_{std}}, \quad w = \frac{(V - V_b)F}{m}, \quad m_{w,c} = \frac{QM_w}{2F_c}\)

Variables

Symbol Meaning SI unit
\(m_{std}\) Water in the standard kg
\(V_{std}\) Titrant used on the standard m³
\(V\) Titrant used on the sample m³
\(V_b\) Blank (drift) volume m³
\(m\) Sample mass kg
\(Q\) Coulometric charge (0 to skip) C
\(F\) Titer (mg water per mL) -
\(m_w\) Water found kg
\(w\) Water content -
\(m_{w,c}\) Water from the coulometric charge kg

Method and standards

Karl Fischer.

Where it is used

  • Moisture in oils, solvents and transformer oil.
  • Pharmaceutical and food water content.

Common mistakes

  • Forgetting the blank.
  • Mixing titer units (mg/mL).

Assumptions and limits

  • Stoichiometric KF reaction; no side reactions.

References

  • Karl Fischer stoichiometry (1 H₂O : 1 I₂ : 2 e⁻)

Questions and answers

Why determine the titer often?

Volumetric reagents lose strength with time and moisture pick-up.

When is coulometric better?

For small amounts of water (ppm levels) because it needs no titer and measures micrograms.

What do the defaults represent?

An illustrative titration: 30 mg water standard uses 6.02 mL; a 2 g oil sample uses 3.45 mL with 0.05 mL blank; 1.2 C coulometric check.

Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Sources

Scholz, Karl Fischer Titration (Springer, 1984), Lipták (ed.), Instrument Engineers' Handbook, Vol. 1: Process Measurement and Analysis, 4th ed. (CRC Press, 2003)

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.