Worked example
Example: Illustrative 50 and 40 mg/L calibration. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Calibration analyte peak area | 15,400 - |
| Calibration analyte concentration | 50.000 mg/L |
| Calibration internal standard area | 12,100 - |
| Calibration internal standard concentration | 40.000 mg/L |
| Sample analyte area | 9,800.0 - |
| Sample internal standard area | 11,500 - |
| Internal standard added to sample | 40.000 mg/L |
Applying the formula below gives:
| Result | Value |
|---|---|
| Relative response factor | 1.0182 |
| Analyte concentration in sample | 33.478 mg/L |
Open this example in the calculator and change any input.
Formula
\(RRF = \frac{A_s/C_s}{A_{is,s}/C_{is,s}}, \quad C_x = \frac{A_x}{A_{is,x}}\frac{C_{is,x}}{RRF}\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(A_{s}\) | Calibration analyte peak area | - |
| \(C_{s}\) | Calibration analyte concentration | kg/m³ |
| \(A_{is,s}\) | Calibration internal standard area | - |
| \(C_{is,s}\) | Calibration internal standard concentration | kg/m³ |
| \(A_x\) | Sample analyte area | - |
| \(A_{is,x}\) | Sample internal standard area | - |
| \(C_{is,x}\) | Internal standard added to sample | kg/m³ |
| \(RRF\) | Relative response factor | - |
| \(C_x\) | Analyte concentration in sample | kg/m³ |
Method and standards
Internal standard.
Where it is used
- GC and HPLC process and laboratory analysis.
- Fuel, solvent and ethanol congener analysis.
Common mistakes
- Using an internal standard present in the sample.
- Adding it at a different stage from calibration.
Assumptions and limits
- Linear detector response through the origin.
References
- Internal standard method
Questions and answers
Why use an internal standard?
Injection volume, evaporation and detector sensitivity errors affect analyte and standard alike and cancel in the area ratio.
Is one calibration point enough?
Only if the response is proven linear through the origin; otherwise use a multi-point calibration of area ratio against concentration ratio.
What do the defaults represent?
Illustrative areas from a 50 mg/L analyte and 40 mg/L internal standard calibration, then a sample spiked with the same 40 mg/L internal standard.
Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.
Sources
Harris, Quantitative Chemical Analysis, 9th ed. (Freeman, 2015)