Calibration Line, Limit of Detection and Quantitation

Least-squares calibration line from three to six standards, the residual standard deviation, limits of detection and quantitation, and a sample's concentration.

Process steam and utilitiesRefinery and petrochemicalChemical and fertiliser

Inputs

Examples: Illustrative six standards 0 to 5 mg/LFour standards, noisyThree standards trace level

Accepted: 3 to 6
Accepted: 0 to 1E+12
Accepted: 0 to 1,000,000 mg/L
Accepted: -1E+12 to 1E+12
Accepted: 0 to 1,000,000 mg/L
Accepted: -1E+12 to 1E+12
Accepted: 0 to 1,000,000 mg/L
Accepted: -1E+12 to 1E+12
Accepted: 0 to 1,000,000 mg/L
Accepted: -1E+12 to 1E+12
Accepted: 0 to 1,000,000 mg/L
Accepted: -1E+12 to 1E+12
Accepted: 0 to 1,000,000 mg/L
Accepted: -1E+12 to 1E+12

Results

y = b + mx by least squares; s_y/x = √(Σe²/(n − 2)); LOD = 3.3s/m and LOQ = 10s/m (the calibration-curve approach used in method validation). Blank replicates give an alternative LOD from the blank standard deviation; report which was used.

Slope197.14
Intercept0.008476
Residual standard deviation0.005667
Limit of detection0.094867mg/L
Limit of quantitation0.28748mg/L
Sample concentration2.0367mg/L

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: Illustrative six standards 0 to 5 mg/L. The inputs, in the units selected above:

InputValue
Number of standards6.0000 -
Sample signal0.41000 -
Standard 10 mg/L
Signal 10.012000 -
Standard 21.0000 mg/L
Signal 20.20500 -
Standard 32.0000 mg/L
Signal 30.39800 -
Standard 43.0000 mg/L
Signal 40.60300 -
Standard 54.0000 mg/L
Signal 50.79000 -
Standard 65.0000 mg/L
Signal 61.0000 -

Applying the formula below gives:

ResultValue
Slope197.14
Intercept0.008476
Residual standard deviation0.005667
Limit of detection0.094867 mg/L
Limit of quantitation0.28748 mg/L
Sample concentration2.0367 mg/L

Open this example in the calculator and change any input.

Formula

\(m = \frac{\sum(x - \bar x)(y - \bar y)}{\sum(x - \bar x)^2}, \quad s_{y/x} = \sqrt{\frac{\sum e^2}{n - 2}}, \quad LOD = \frac{3.3s_{y/x}}{m}, \quad LOQ = \frac{10s_{y/x}}{m}\)

Variables

Symbol Meaning SI unit
\(n\) Number of standards -
\(y_x\) Sample signal -
\(x_1\) Standard 1 kg/m³
\(y_1\) Signal 1 -
\(x_2\) Standard 2 kg/m³
\(y_2\) Signal 2 -
\(x_3\) Standard 3 kg/m³
\(y_3\) Signal 3 -
\(x_4\) Standard 4 kg/m³
\(y_4\) Signal 4 -
\(x_5\) Standard 5 kg/m³
\(y_5\) Signal 5 -
\(x_6\) Standard 6 kg/m³
\(y_6\) Signal 6 -
\(m\) Slope -
\(b\) Intercept -
\(s_{y/x}\) Residual standard deviation -
\(LOD\) Limit of detection kg/m³
\(LOQ\) Limit of quantitation kg/m³
\(C_x\) Sample concentration kg/m³

Method and standards

Calibration statistics.

Where it is used

  • Analyser and laboratory method validation.
  • Checking whether a result is reportable.

Common mistakes

  • Extrapolating beyond the highest standard.
  • Reporting results below the LOQ as exact numbers.

Assumptions and limits

  • Linear response; constant variance.

References

  • Calibration-curve LOD and LOQ (3.3σ/S and 10σ/S)

Questions and answers

What do LOD and LOQ mean?

LOD is the smallest concentration distinguishable from zero; LOQ the smallest measured with acceptable precision, about three times larger.

Why n − 2 in the standard deviation?

Two degrees of freedom are used to fit the slope and intercept.

What do the defaults represent?

Six illustrative standards from 0 to 5 mg/L with near-linear signals and a sample reading 0.41.

Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Sources

Harris, Quantitative Chemical Analysis, 9th ed. (Freeman, 2015)

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.