Worked example
Example: Illustrative six standards 0 to 5 mg/L. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Number of standards | 6.0000 - |
| Sample signal | 0.41000 - |
| Standard 1 | 0 mg/L |
| Signal 1 | 0.012000 - |
| Standard 2 | 1.0000 mg/L |
| Signal 2 | 0.20500 - |
| Standard 3 | 2.0000 mg/L |
| Signal 3 | 0.39800 - |
| Standard 4 | 3.0000 mg/L |
| Signal 4 | 0.60300 - |
| Standard 5 | 4.0000 mg/L |
| Signal 5 | 0.79000 - |
| Standard 6 | 5.0000 mg/L |
| Signal 6 | 1.0000 - |
Applying the formula below gives:
| Result | Value |
|---|---|
| Slope | 197.14 |
| Intercept | 0.008476 |
| Residual standard deviation | 0.005667 |
| Limit of detection | 0.094867 mg/L |
| Limit of quantitation | 0.28748 mg/L |
| Sample concentration | 2.0367 mg/L |
Open this example in the calculator and change any input.
Formula
\(m = \frac{\sum(x - \bar x)(y - \bar y)}{\sum(x - \bar x)^2}, \quad s_{y/x} = \sqrt{\frac{\sum e^2}{n - 2}}, \quad LOD = \frac{3.3s_{y/x}}{m}, \quad LOQ = \frac{10s_{y/x}}{m}\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(n\) | Number of standards | - |
| \(y_x\) | Sample signal | - |
| \(x_1\) | Standard 1 | kg/m³ |
| \(y_1\) | Signal 1 | - |
| \(x_2\) | Standard 2 | kg/m³ |
| \(y_2\) | Signal 2 | - |
| \(x_3\) | Standard 3 | kg/m³ |
| \(y_3\) | Signal 3 | - |
| \(x_4\) | Standard 4 | kg/m³ |
| \(y_4\) | Signal 4 | - |
| \(x_5\) | Standard 5 | kg/m³ |
| \(y_5\) | Signal 5 | - |
| \(x_6\) | Standard 6 | kg/m³ |
| \(y_6\) | Signal 6 | - |
| \(m\) | Slope | - |
| \(b\) | Intercept | - |
| \(s_{y/x}\) | Residual standard deviation | - |
| \(LOD\) | Limit of detection | kg/m³ |
| \(LOQ\) | Limit of quantitation | kg/m³ |
| \(C_x\) | Sample concentration | kg/m³ |
Method and standards
Calibration statistics.
Where it is used
- Analyser and laboratory method validation.
- Checking whether a result is reportable.
Common mistakes
- Extrapolating beyond the highest standard.
- Reporting results below the LOQ as exact numbers.
Assumptions and limits
- Linear response; constant variance.
References
- Calibration-curve LOD and LOQ (3.3σ/S and 10σ/S)
Questions and answers
What do LOD and LOQ mean?
LOD is the smallest concentration distinguishable from zero; LOQ the smallest measured with acceptable precision, about three times larger.
Why n − 2 in the standard deviation?
Two degrees of freedom are used to fit the slope and intercept.
What do the defaults represent?
Six illustrative standards from 0 to 5 mg/L with near-linear signals and a sample reading 0.41.
Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.
Sources
Harris, Quantitative Chemical Analysis, 9th ed. (Freeman, 2015)