Worked example
Example: Soda-lime hydrometer 951.0 at 25 °C. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Reading (density graduated at 20 °C) | 951.00 kg/m³ |
| Temperature of the liquid | 25.000 °C |
| Instrument glass | Soda-lime glass (25 × 10⁻⁶ /°C) |
Applying the formula below gives:
| Result | Value |
|---|---|
| Cubical expansion used | 25.000 µm/(m·K) |
| True density at t | 950.88 kg/m³ |
| Strength by mass | 29.859 % |
| Strength by volume at 20 °C | 36.093 % |
Open this example in the calculator and change any input.
Formula
\(\rho_t = \rho'[1 - a(t - 20)], \quad \rho(p, t) = \rho_t\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(\rho'\) | Reading (density graduated at 20 °C) | kg/m³ |
| \(t\) | Temperature of the liquid | K |
| \(\text{glass}\) | Instrument glass | - |
| \(a\) | Cubical expansion used | 1/K |
| \(\rho_t\) | True density at t | kg/m³ |
| \(p\) | Strength by mass | - |
| \(q\) | Strength by volume at 20 °C | - |
Method and standards
R 22 Annex I method for Tables IX and X.
Where it is used
- Bonded warehouse gauging.
- Quality control of spirits.
Common mistakes
- Using the reading directly as the density at t.
- Wrong glass coefficient.
Assumptions and limits
- Graduated at 20 °C; no surface-tension correction.
References
- OIML R 22 (1973)
- Alcoholometry
- and International Alcoholometric Tables (1975)
- Annex I
- Tables IXa
- IXb
- Xa and Xb
Questions and answers
What is a hydrometer for alcohol?
A glass hydrometer graduated in density at 20 °C rather than in strength.
Why two glasses?
R 22 uses 25 × 10⁻⁶ /°C for soda-lime glass and 10 × 10⁻⁶ /°C for borosilicate.
Does an electronic meter need this?
No, oscillating-tube meters compensate internally.
Method and article reviewed by FireCAD engineering on 13 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.