Ultrasonic Thickness Gauge: Velocity Calibration and Wall Loss

Sound velocity from a reference block, wall thickness from the pulse-echo time with temperature correction, and the wall loss against nominal thickness.

Inputs

Examples: Illustrative hot steel pipeAmbient tank shell, no correctionAluminium vessel

Accepted: 0.1 to 1,000 mm
Accepted: 0.01 to 1,000 µs
Accepted: 0.01 to 1,000 µs
Accepted: -0.01 to 0
Accepted: -100 to 500 K
Accepted: 0.1 to 1,000 mm

Results

Pulse-echo: the sound crosses the wall twice, so t = vΔt/2. Velocity is set on a block of the same material and known thickness. Sound slows in hot steel (about 1 % per 100 K for longitudinal waves), so a gauge calibrated cold over-reads hot walls unless corrected. Coatings, curvature and pitting affect readings; use dual-element probes for corroded surfaces.

Calibrated sound velocity6,349.2m/s
Thickness without temperature correction5.7778mm
Thickness corrected for temperature5.7316mm
Wall loss0.26844mm
Wall loss share4.4741%

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: Illustrative hot steel pipe. The inputs, in the units selected above:

InputValue
Calibration block thickness20.000 mm
Round-trip time in the block6.7800 µs
Round-trip time on the part3.0500 µs
Velocity temperature coefficient-1.0000e-04 -
Part temperature above calibration150.00 K
Nominal wall thickness10.200 mm

Applying the formula below gives:

ResultValue
Calibrated sound velocity5,899.7 m/s
Thickness without temperature correction8.9971 mm
Thickness corrected for temperature8.8621 mm
Wall loss1.3379 mm
Wall loss share13.117 %

Open this example in the calculator and change any input.

Formula

\(v = \frac{2t_{ref}}{\Delta t_{ref}}, \quad t = \frac{v\Delta t}{2}(1 + k_T\Delta T)\)

Variables

Symbol Meaning SI unit
\(t_{ref}\) Calibration block thickness m
\(\Delta t_{ref}\) Round-trip time in the block s
\(\Delta t\) Round-trip time on the part s
\(k_T\) Velocity temperature coefficient -
\(T - T_{cal}\) Part temperature above calibration K
\(t_n\) Nominal wall thickness m
\(v\) Calibrated sound velocity m/s
\(t_{uncorr}\) Thickness without temperature correction m
\(t\) Thickness corrected for temperature m
\(t_n - t\) Wall loss m
\((t_n - t)/t_n\) Wall loss share -

Method and standards

UT thickness.

Where it is used

  • Corrosion monitoring of pipes and vessels.
  • Remaining-life inspection.

Common mistakes

  • Calibrating on a different alloy.
  • Ignoring temperature on hot lines.

Assumptions and limits

  • Longitudinal waves; parallel surfaces; linear velocity–temperature relation.

References

  • Pulse-echo thickness measurement

Questions and answers

Why calibrate on a block?

Sound velocity varies with alloy and heat treatment; a known block of the same material sets it.

Does temperature matter?

Yes: sound slows in hot steel, so a gauge calibrated cold reads a hot wall as thicker than it is unless corrected.

What do the defaults represent?

An illustrative steel check: 20 mm block at 6.78 µs, part at 3.05 µs and 150 K hotter, nominal 10.2 mm wall.

Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Sources

Krautkrämer and Krautkrämer, Ultrasonic Testing of Materials, 4th ed. (Springer, 1990), Lipták (ed.), Instrument Engineers' Handbook, Vol. 1: Process Measurement and Analysis, 4th ed. (CRC Press, 2003)

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.