Worked example
Example: Illustrative 10 pC/g sensor at 5 g. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Charge sensitivity (pC per g) | 10.000 - |
| Acceleration | 5.0000 g |
| Charge amplifier feedback capacitance | 1,000.0 pF |
| Feedback resistance | 1,000.0 MΩ |
| Accelerometer capacitance | 1,200.0 pF |
| Cable capacitance | 1,000.0 pF |
Applying the formula below gives:
| Result | Value |
|---|---|
| Charge | 50.000 pC |
| Charge amplifier output | 50.000 mV |
| Voltage-mode output with cable | 22.727 mV |
| Voltage loss from the cable | 45.455 % |
| Low-frequency cut-off | 0.15915 Hz |
Open this example in the calculator and change any input.
Formula
\(q = S_qa, \quad V_o = \frac{q}{C_f}, \quad V_v = \frac{q}{C_s + C_c}, \quad f_L = \frac1{2\pi R_fC_f}\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(S_q\) | Charge sensitivity (pC per g) | - |
| \(a\) | Acceleration | m/s² |
| \(C_f\) | Charge amplifier feedback capacitance | F |
| \(R_f\) | Feedback resistance | Ω |
| \(C_s\) | Accelerometer capacitance | F |
| \(C_c\) | Cable capacitance | F |
| \(q\) | Charge | C |
| \(V_o\) | Charge amplifier output | V |
| \(V_v\) | Voltage-mode output with cable | V |
| \(C_c/(C_s + C_c)\) | Voltage loss from the cable | - |
| \(f_L\) | Low-frequency cut-off | 1/s |
Method and standards
Charge amplifier.
Where it is used
- Vibration measurement chain design.
- Troubleshooting low readings.
Common mistakes
- Forgetting cable capacitance in voltage mode.
- Mixing pC/g and pC/(m/s²).
Assumptions and limits
- Ideal charge amplifier; linear sensor.
References
- Piezoelectric transducer and charge amplifier
Questions and answers
Why use a charge amplifier?
Its output does not depend on cable capacitance, so long or changing cables do not alter the calibration.
What limits low-frequency measurement?
Charge leaks away through the feedback resistor; the time constant R_fC_f sets the lowest usable frequency.
What do the defaults represent?
An illustrative 10 pC/g accelerometer (1.2 nF) at 5 g with a 1 nF feedback capacitor, 1 GΩ feedback resistor and 1 nF of cable.
Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.