Worked example
Example: Illustrative pH 7 and 4 buffers at 25 C. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Buffer 1 pH | 7.0000 - |
| Reading in buffer 1 | -3.0000 mV |
| Buffer 2 pH | 4.0100 - |
| Reading in buffer 2 | 166.50 mV |
| Calibration temperature | 25.000 °C |
| Sample reading | 50.000 mV |
Applying the formula below gives:
| Result | Value |
|---|---|
| Electrode slope | 56.689 mV |
| Theoretical Nernst slope | 59.159 mV |
| Slope efficiency | 95.824 % |
| Offset (reading at pH 7) | -3.0000 mV |
| Sample pH | 6.0651 |
Open this example in the calculator and change any input.
Formula
\(s = \frac{E_2 - E_1}{pH_1 - pH_2}, \quad s_N = \frac{\ln10\,RT}{F}, \quad pH_x = 7 - \frac{E_x - E_{pH7}}{s}\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(pH_1\) | Buffer 1 pH | - |
| \(E_1\) | Reading in buffer 1 | V |
| \(pH_2\) | Buffer 2 pH | - |
| \(E_2\) | Reading in buffer 2 | V |
| \(T\) | Calibration temperature | K |
| \(E_x\) | Sample reading | V |
| \(s\) | Electrode slope | V |
| \(s_N\) | Theoretical Nernst slope | V |
| \(s/s_N\) | Slope efficiency | - |
| \(E_{pH7}\) | Offset (reading at pH 7) | V |
| \(pH_x\) | Sample pH | - |
Method and standards
pH calibration.
Where it is used
- Routine analyser and laboratory meter calibration.
- Electrode health checks.
Common mistakes
- Calibrating at one temperature and measuring at another.
- Using expired or contaminated buffers.
Assumptions and limits
- Linear electrode response between buffers; buffers at stated pH for the temperature.
References
- Two-point calibration of a glass electrode
Questions and answers
What slope is acceptable?
Typically 95–102 % of the Nernst value; below about 90 % clean or replace the electrode.
Why bracket the sample with buffers?
The electrode is closest to linear between the buffers; extrapolating far outside them adds error.
What do the defaults represent?
An illustrative calibration at 25 °C with pH 7.00 and 4.01 buffers reading −3 and +166.5 mV, and a sample at +50 mV.
Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.