ORP (Redox Potential) by the Nernst Equation

Redox potential of a couple from its standard potential, oxidised and reduced activities, pH and temperature, versus hydrogen and Ag/AgCl references.

Inputs

Examples: Illustrative Fe3+/Fe2+ at pH 2Hypochlorite/chloride at pH 7Dichromate/Cr3+ at pH 2.5, 40 C

Accepted: -3,000 to 3,000 mV
Accepted: 1 to 8
Accepted: 0 to 16
Accepted: 1E-12 to 10
Accepted: 1E-12 to 10
Accepted: 0 to 14
Accepted: 0 to 100 °C
Accepted: -1,000 to 1,000 mV

Results

For Ox + mH⁺ + ne⁻ ⇌ Red: E_h = E⁰ + (2.303RT/nF)log(a_ox/a_red) − (2.303RT·m/nF)pH. ORP meters read against a reference electrode, usually Ag/AgCl in saturated KCl (+0.197 V vs SHE at 25 °C), so E_meas = E_h − E_ref. Real ORP readings in mixed waters are often not at equilibrium and are best used as trends (for example cyanide destruction, chlorination, chromate reduction).

Slope per decade10.356mV
Potential vs SHE977.90mV
Reading vs the reference electrode789.90mV

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: Illustrative Fe3+/Fe2+ at pH 2. The inputs, in the units selected above:

InputValue
Standard potential vs SHE771.00 mV
Electrons transferred1.0000 -
Protons consumed in the reduction0 -
Oxidised species activity0.001000 -
Reduced species activity1.0000e-04 -
pH2.0000 -
Temperature25.000 °C
Reference electrode potential vs SHE197.00 mV

Applying the formula below gives:

ResultValue
Slope per decade59.159 mV
Potential vs SHE830.16 mV
Reading vs the reference electrode633.16 mV

Open this example in the calculator and change any input.

Formula

\(E_h = E^0 + \frac{2.303RT}{nF}\log\frac{a_{ox}}{a_{red}} - \frac{2.303RT\,m}{nF}pH, \quad E_{meas} = E_h - E_{ref}\)

Variables

Symbol Meaning SI unit
\(E^0\) Standard potential vs SHE V
\(n\) Electrons transferred -
\(m\) Protons consumed in the reduction -
\(a_{ox}\) Oxidised species activity -
\(a_{red}\) Reduced species activity -
\(pH\) pH -
\(T\) Temperature K
\(E_{ref}\) Reference electrode potential vs SHE V
\(S\) Slope per decade V
\(E_h\) Potential vs SHE V
\(E_{meas}\) Reading vs the reference electrode V

Method and standards

ORP.

Where it is used

  • Cyanide, chrome and chlorination control.
  • Water quality monitoring.

Common mistakes

  • Comparing Ag/AgCl readings with SHE values.
  • Treating ORP as a concentration measurement.

Assumptions and limits

  • Equilibrium; activities known; reversible couple.

References

  • Nernst equation for a redox couple

Questions and answers

Why does my ORP reading differ from the table value?

Meters read against Ag/AgCl, about 200 mV below the hydrogen electrode, and real solutions are rarely at equilibrium.

Does pH affect ORP?

Yes, for reactions involving hydrogen ions: about 59 mV × m/n per pH unit at 25 °C.

What do the defaults represent?

The iron(III)/iron(II) couple (E⁰ 0.771 V) with a 10:1 ratio at pH 2 and 25 °C, read against Ag/AgCl (0.197 V).

Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Sources

Bard and Faulkner, Electrochemical Methods, 2nd ed. (Wiley, 2001), Lipták (ed.), Instrument Engineers' Handbook, Vol. 1: Process Measurement and Analysis, 4th ed. (CRC Press, 2003)

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.