Ion-Selective Electrode Interference (Nikolsky–Eisenman)

Primary ion activity from an ion-selective electrode reading with the Nikolsky–Eisenman correction for an interfering ion, and the error if it is ignored.

Inputs

Examples: Illustrative sodium electrode with potassiumCalcium electrode with magnesiumNitrate electrode with chloride

Accepted: -2,000 to 2,000 mV
Accepted: -2,000 to 2,000 mV
Accepted: -3 to 3
Accepted: -3 to 3
Accepted: 0 to 10
Accepted: 0 to 10
Accepted: 0 to 100 °C

Results

Nikolsky–Eisenman: E = E⁰ + (2.303RT/z_iF)·log(a_i + K_ij·a_j^{z_i/z_j}). The electrode sees the interfering ion as K_ij times as much primary ion; with K_ij from the maker or a separate-solution test, the true activity is the apparent one minus K_ij·a_j^{z_i/z_j}. Activities, not concentrations: use an ionic-strength adjustment buffer so the activity coefficient is constant. Slope 59.16 mV/decade at 25 °C for z = 1 (negative for anions).

Nernst slope per decade59.159mV
Apparent activity (no correction, mol/L)0.001019
Primary ion activity (mol/L)5.1878e-04
Error if interference ignored96.380%

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: Illustrative sodium electrode with potassium. The inputs, in the units selected above:

InputValue
Measured potential118.00 mV
Electrode intercept (from calibration)295.00 mV
Charge of the primary ion1.0000 -
Charge of the interfering ion1.0000 -
Selectivity coefficient0.010000 -
Interfering ion activity (mol/L)0.050000 -
Temperature25.000 °C

Applying the formula below gives:

ResultValue
Nernst slope per decade59.159 mV
Apparent activity (no correction, mol/L)0.001019
Primary ion activity (mol/L)5.1878e-04
Error if interference ignored96.380 %

Open this example in the calculator and change any input.

Formula

\(E = E^0 + \frac{2.303RT}{z_iF}\log\left(a_i + K_{ij}a_j^{z_i/z_j}\right)\)

Variables

Symbol Meaning SI unit
\(E\) Measured potential V
\(E^0\) Electrode intercept (from calibration) V
\(z_i\) Charge of the primary ion -
\(z_j\) Charge of the interfering ion -
\(K_{ij}\) Selectivity coefficient -
\(a_j\) Interfering ion activity (mol/L) -
\(T\) Temperature K
\(S\) Nernst slope per decade V
\(a_{app}\) Apparent activity (no correction, mol/L) -
\(a_i\) Primary ion activity (mol/L) -
\(a_{app}/a_i - 1\) Error if interference ignored -

Method and standards

ISE interference.

Where it is used

  • Water and process analysers with ion-selective electrodes.
  • Choosing electrodes for a matrix.

Common mistakes

  • Ignoring interference in mixed matrices.
  • Confusing activity and concentration.

Assumptions and limits

  • Nernstian electrode; fixed ionic strength; known selectivity coefficient.

References

  • Nikolsky–Eisenman equation

Questions and answers

What is the selectivity coefficient?

How strongly the electrode responds to the interfering ion relative to the primary ion; smaller is better.

Why use an ionic strength buffer?

Electrodes respond to activity; a buffer fixes the activity coefficient so readings track concentration.

What do the defaults represent?

An illustrative sodium electrode with E⁰ 295.0 mV reading 118.0 mV, with 0.05 mol/L potassium at K = 0.01, 25 °C.

Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Sources

Buck and Lindner, Recommendations for Nomenclature of Ion-Selective Electrodes, Pure Appl. Chem. 66 (1994), Bard and Faulkner, Electrochemical Methods, 2nd ed. (Wiley, 2001)

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.