Beer–Lambert Absorbance and Concentration

Absorbance from measured transmittance and the analyte concentration from molar absorptivity and path length, in molar and mass units.

Process steam and utilitiesRefinery and petrochemicalChemical and fertiliser

Inputs

Examples: Illustrative NADH, 35 % TNitrate UV, 1 mm cellDye, 5 cm process cell

Accepted: 1E-4 to 100 %
Accepted: 0.001 to 1,000,000
Accepted: 0.001 to 1,000 cm
Accepted: 1 to 1,000,000 kg/kmol

Results

A = −log₁₀T = εlc. ε in L/(mol·cm) equals 0.1 m²/mol. Keep absorbance roughly between 0.1 and 1 for best precision; above about 1–2 stray light and chemical effects bend the calibration line. Calibrate with standards rather than trusting a literature ε for routine work.

Absorbance0.22185
Molar concentration0.31693mmol/L
Mass concentration19.649mg/L

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: Illustrative NADH, 35 % T. The inputs, in the units selected above:

InputValue
Transmittance35.000 %
Molar absorptivity (L/(mol·cm))6,220.0 -
Path length1.0000 cm
Molar mass663.40 kg/kmol

Applying the formula below gives:

ResultValue
Absorbance0.45593
Molar concentration0.073301 mmol/L
Mass concentration48.628 mg/L

Open this example in the calculator and change any input.

Formula

\(A = -\log_{10}T = \varepsilon lc, \quad c = \frac{A}{\varepsilon l}\)

Variables

Symbol Meaning SI unit
\(T\) Transmittance -
\(\varepsilon\) Molar absorptivity (L/(mol·cm)) -
\(l\) Path length m
\(M\) Molar mass kg/kmol
\(A\) Absorbance -
\(c\) Molar concentration mol/m³
\(c_m\) Mass concentration kg/m³

Method and standards

Absorbance.

Where it is used

  • Process photometers and colour analysers.
  • Laboratory UV-visible assays.

Common mistakes

  • Using percent transmittance in the logarithm.
  • Ignoring the blank.

Assumptions and limits

  • Monochromatic light; dilute, non-scattering sample.

References

  • Beer–Lambert–Bouguer law

Questions and answers

Why does the calibration curve bend at high absorbance?

Stray light, polychromatic light and molecular interactions break the linear law.

What is a typical path length?

1 cm cuvettes in the laboratory; process photometers use from millimetres to tens of centimetres.

What do the defaults represent?

An illustrative NADH-type measurement: 35 % transmittance in a 1 cm cell with ε = 6,220 L/(mol·cm) and M = 663.4 g/mol.

Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Sources

Skoog, Holler and Crouch, Principles of Instrumental Analysis, 7th ed. (Cengage, 2017)

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.