True alcoholic strength at 20 °C from the reading of a soda-lime glass alcoholometer (% vol or % mass) taken at another temperature, OIML R 22 Tables VIII.

Distillery and ethanol

Inputs

Examples: Reads 42 % vol at 28 °CReads 40 % mass at 10 °C

R 22 formula range −20 to +40 °C.
R 22: soda-lime glass 25 × 10⁻⁶ /°C, the cubical (volume) coefficient, not the linear one (about a third of it).

Results

The reading is also affected by surface tension at temperatures far from 20 °C: see the surface-tension correction.

Density corresponding to the reading935.15kg/m³
True density at t935.38kg/m³
Strength by mass43.498%
Strength by volume at 20 °C51.140%

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: Reads 42 % vol at 28 °C. The inputs, in the units selected above:

InputValue
Alcoholometer graduated in% vol
Reading of the alcoholometer42.000 %
Temperature of the liquid28.000 °C
Cubical (volume) expansion of the glass, about 3 × linear25.000 µm/(m·K)

Applying the formula below gives:

ResultValue
Density corresponding to the reading944.76 kg/m³
True density at t944.57 kg/m³
Strength by mass32.251 %
Strength by volume at 20 °C38.817 %

Open this example in the calculator and change any input.

Formula

\(\rho'_{20} = \rho_{20}(r), \quad \rho_t = \rho'_{20}[1 - a(t - 20)], \quad \rho(p, t) = \rho_t\)

Variables

Symbol Meaning SI unit
\(\text{scale}\) Alcoholometer graduated in -
\(r\) Reading of the alcoholometer -
\(t\) Temperature of the liquid K
\(a\) Cubical (volume) expansion of the glass, about 3 × linear 1/K
\(\rho'_{20}\) Density corresponding to the reading kg/m³
\(\rho_t\) True density at t kg/m³
\(p\) Strength by mass -
\(q\) Strength by volume at 20 °C -

Method and standards

R 22 Annex I method for Tables VIII.

Where it is used

  • Gauging spirits in bonded warehouses.
  • Distillery receiver checks.

Common mistakes

  • Applying a borosilicate coefficient to a soda-lime instrument.
  • Reading the meniscus top instead of the plane of the liquid.

Assumptions and limits

  • Instrument graduated at 20 °C.
  • No surface-tension correction.

References

  • OIML R 22 (1973)
  • Alcoholometry
  • and International Alcoholometric Tables (1975)
  • Annex I
  • Tables VIIIa and VIIIb

Questions and answers

Why does a reading at 30 °C overstate the strength?

Warm spirit is less dense, so the alcoholometer sinks deeper and reads high.

Why the glass expansion?

The stem's volume grows with temperature; R 22 corrects with ρt = ρ'20[1 − a(t − 20)], a = 25 × 10⁻⁶ /°C.

Which tables does this replace?

R 22 Tables VIIIa (% mass) and VIIIb (% vol).

Method and article reviewed by FireCAD engineering on 13 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.