FireCADCalc

Freeze-Drying (Lyophilisation) Primary Drying Time

Sublimation rate and primary drying time of a vial in a freeze dryer from the shelf and product temperatures, the vial heat transfer coefficient and the ice mass.

Inputs

Examples: Illustrative 3 mL vialLarge 10 mL fillConservative cycle

Accepted: 0.1 to 1,000 mL
Accepted: 0 to 50 %
Accepted: 900 to 1,300 kg/m³
Accepted: 0.1 to 100 cm²
Accepted: 1 to 100 W/(m²·K)
Accepted: -73.15 to 56.85 °C
Accepted: -73.15 to -0.15 °C
Accepted: 2,000 to 3,000 kJ/kg

Results

During primary drying the heat that reaches the vial from the shelf, K_v A_v (T_s − T_p), is used to sublime ice, so the sublimation rate is that heat over the heat of sublimation and the drying time is the ice mass over the rate. The product temperature must stay below its collapse (or eutectic) temperature, which limits how warm the shelf can be; K_v depends on the vial and the chamber pressure (about 10–30 W/(m²·K) for glass tubing vials at 10–20 Pa). The dry layer's resistance, ignored here, sets the product temperature: confirm with a cycle run.

Ice per vial9.5680g
Heat flow to the vial383.40mW
Sublimation rate per vial4.8634e-04kg/h
Primary drying time19.673h

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: Illustrative 3 mL vial. The inputs, in the units selected above:

InputValue
Fill volume per vial3.0000 mL
Solids share of the fill5.0000 %
Density of the solution1,030.0 kg/m³
Vial bottom area (outer)3.8000 cm²
Vial heat transfer coefficient20.000 W/(m²·K)
Shelf temperature-10.000 °C
Product temperature at the sublimation front-30.000 °C
Heat of sublimation of ice2,838.0 kJ/kg

Applying the formula below gives:

ResultValue
Ice per vial2.9355 g
Heat flow to the vial152.00 mW
Sublimation rate per vial1.9281e-04 kg/h
Primary drying time15.225 h

Open this example in the calculator and change any input.

Formula

\(\dot m = \frac{K_v A_v (T_s - T_p)}{\Delta H_s}, \quad t = \frac{m_{ice}}{\dot m}\)

Variables

Symbol Meaning SI unit
\(V_f\) Fill volume per vial m³
\(x_s\) Solids share of the fill -
\(\rho\) Density of the solution kg/m³
\(A_v\) Vial bottom area (outer) m²
\(K_v\) Vial heat transfer coefficient W/(m²·K)
\(T_s\) Shelf temperature K
\(T_p\) Product temperature at the sublimation front K
\(\Delta H_s\) Heat of sublimation of ice J/kg
\(m_{ice}\) Ice per vial kg
\(\dot Q\) Heat flow to the vial W
\(\dot m\) Sublimation rate per vial kg/s
\(t\) Primary drying time s

Method and standards

Primary drying.

Where it is used

  • First estimates of lyophilisation cycles.
  • Comparing vial types and shelf temperatures.

Common mistakes

  • Using the chamber pressure-independent K_v.
  • Forgetting that the solids do not sublime.

Assumptions and limits

  • Heat-transfer-limited drying; uniform vials; constant temperatures.

References

  • Steady-state heat-transfer model of primary drying (vial heat transfer coefficient)

Questions and answers

How long does primary drying take?

For the defaults, 3 mL at 5 % solids in a 22 mm vial with K_v = 20 W/(m²·K) and 20 K driving force, about 15 h.

What limits the shelf temperature?

The product must stay below its collapse temperature at the sublimation front.

What do the defaults represent?

An illustrative 3 mL fill at 5 % solids in a vial of 3.8 cm² base, shelf −10 °C, product −30 °C, K_v 20 W/(m²·K).

Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Sources

Pikal, Freeze-drying of proteins: process, formulation and stability (ACS Symposium Series 567, 1994)

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.