Worked example
Example: Illustrative 3 mL vial. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Fill volume per vial | 3.0000 mL |
| Solids share of the fill | 5.0000 % |
| Density of the solution | 1,030.0 kg/m³ |
| Vial bottom area (outer) | 3.8000 cm² |
| Vial heat transfer coefficient | 20.000 W/(m²·K) |
| Shelf temperature | -10.000 °C |
| Product temperature at the sublimation front | -30.000 °C |
| Heat of sublimation of ice | 2,838.0 kJ/kg |
Applying the formula below gives:
| Result | Value |
|---|---|
| Ice per vial | 2.9355 g |
| Heat flow to the vial | 152.00 mW |
| Sublimation rate per vial | 1.9281e-04 kg/h |
| Primary drying time | 15.225 h |
Open this example in the calculator and change any input.
Formula
\(\dot m = \frac{K_v A_v (T_s - T_p)}{\Delta H_s}, \quad t = \frac{m_{ice}}{\dot m}\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(V_f\) | Fill volume per vial | m³ |
| \(x_s\) | Solids share of the fill | - |
| \(\rho\) | Density of the solution | kg/m³ |
| \(A_v\) | Vial bottom area (outer) | m² |
| \(K_v\) | Vial heat transfer coefficient | W/(m²·K) |
| \(T_s\) | Shelf temperature | K |
| \(T_p\) | Product temperature at the sublimation front | K |
| \(\Delta H_s\) | Heat of sublimation of ice | J/kg |
| \(m_{ice}\) | Ice per vial | kg |
| \(\dot Q\) | Heat flow to the vial | W |
| \(\dot m\) | Sublimation rate per vial | kg/s |
| \(t\) | Primary drying time | s |
Method and standards
Primary drying.
Where it is used
- First estimates of lyophilisation cycles.
- Comparing vial types and shelf temperatures.
Common mistakes
- Using the chamber pressure-independent K_v.
- Forgetting that the solids do not sublime.
Assumptions and limits
- Heat-transfer-limited drying; uniform vials; constant temperatures.
References
- Steady-state heat-transfer model of primary drying (vial heat transfer coefficient)
Questions and answers
How long does primary drying take?
For the defaults, 3 mL at 5 % solids in a 22 mm vial with K_v = 20 W/(m²·K) and 20 K driving force, about 15 h.
What limits the shelf temperature?
The product must stay below its collapse temperature at the sublimation front.
What do the defaults represent?
An illustrative 3 mL fill at 5 % solids in a vial of 3.8 cm² base, shelf −10 °C, product −30 °C, K_v 20 W/(m²·K).
Method and article reviewed by FireCAD engineering on 17 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.
Sources
Pikal, Freeze-drying of proteins: process, formulation and stability (ACS Symposium Series 567, 1994)