FireCADCalc

Dead-Weight Tester Pressure with Local Gravity and Air Buoyancy

Pressure from a dead-weight tester from mass, piston area, local gravity and air buoyancy, and the error made by assuming standard gravity.

Thermal power plantProcess steam and utilitiesSugar mill and cogenerationDistillery and ethanolRefinery and petrochemicalChemical and fertiliser

Inputs

Examples: 10 kg on 1 cm², g = 9.786650 kg on 0.1 cm², g = 9.7833

Results

Gravity in India ranges from about 9.776 to 9.790 m/s²; using standard gravity overstates pressure by up to 0.3 %. Piston area also changes with temperature and pressure (makers' coefficients).

Generated pressure9.7851bar
Pressure assuming 9.80665 m/s² and no buoyancy9.8066bar
Error of the simple calculation0.22006%

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: 10 kg on 1 cm², g = 9.7866. The inputs, in the units selected above:

InputValue
Mass of weights and piston10.000 kg
Effective piston area1.0000 cm²
Local gravity9.7866 m/s²
Air density1.2000 kg/m³
Density of the weights7,920.0 kg/m³

Applying the formula below gives:

ResultValue
Generated pressure9.7851 bar
Pressure assuming 9.80665 m/s² and no buoyancy9.8066 bar
Error of the simple calculation0.22006 %

Open this example in the calculator and change any input.

Formula

\(p = \frac{mg_l(1 - \rho_a/\rho_m)}{A_e}\)

Variables

Symbol Meaning SI unit
\(m\) Mass of weights and piston kg
\(A_e\) Effective piston area m²
\(g_l\) Local gravity, in m/s² m/s²
\(\rho_a\) Air density kg/m³
\(\rho_m\) Density of the weights kg/m³
\(p\) Generated pressure Pa
\(p_{std}\) Pressure assuming 9.80665 m/s² and no buoyancy Pa
\(\Delta p/p\) Error of the simple calculation -

Method and standards

Pressure balance.

Where it is used

  • Pressure gauge and transmitter calibration.
  • Laboratory accuracy checks.

Common mistakes

  • Using standard gravity.
  • Ignoring piston temperature.

Assumptions and limits

  • Effective area at reference conditions: the makers' temperature and pressure-distortion coefficients for the piston are not applied.
  • No fluid-head correction between the piston reference level and the device under test.
  • Local gravity is entered in m/s² in every unit system.

References

  • Pressure balance p = mg(1 − ρ_a/ρ_m)/A_e (calibration practice; OIML R 110 principle)

Questions and answers

Why does local gravity matter?

Weights press with mg; gravity varies about 0.5 % between the equator and poles and with altitude.

Why air buoyancy?

Air displaced by the weights pushes up with about 0.015 % of their weight for steel weights.

What do the defaults represent?

10 kg on a 1 cm² piston at a gravity typical of central India.

Method and article reviewed by FireCAD engineering on 13 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.