Worked example
Example: 2 500 µS/cm at 45 °C, 2 %/K. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Measured conductivity | 2,500.0 - |
| Sample temperature | 45.000 °C |
| Temperature coefficient (per K) | 0.020000 - |
| TDS factor (mg/L per µS/cm) | 0.65000 - |
Applying the formula below gives:
| Result | Value |
|---|---|
| Conductivity at 25 °C | 1,785.7 |
| Ratio of measured to reference | 1.4000 |
| Estimated TDS (mg/L) | 1,160.7 |
Open this example in the calculator and change any input.
Formula
\(\kappa_{25} = \frac{\kappa_T}{1 + \alpha (T - 25\ °C)}, \quad TDS = f\, \kappa_{25}\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(\kappa_T\) | Measured conductivity | - |
| \(T\) | Sample temperature | K |
| \(\alpha\) | Temperature coefficient (per K) | - |
| \(f\) | TDS factor (mg/L per µS/cm) | - |
| \(\kappa_{25}\) | Conductivity at 25 °C | - |
| \(\kappa_T/\kappa_{25}\) | Ratio of measured to reference | - |
| \(TDS\) | Estimated TDS (mg/L) | - |
Method and standards
Linear compensation.
Where it is used
- Boiler water and cooling water monitoring.
- Effluent testing.
Common mistakes
- Comparing uncompensated readings with limits.
- One TDS factor for all waters.
Assumptions and limits
- Linear coefficient over the range. High-purity water (below about 20 µS/cm) and samples more than 25 K from 25 °C are flagged: their temperature dependence is non-linear.
- Conductivity in µS/cm, the basis of the TDS factor.
References
- Linear temperature compensation κ₂₅ = κ_T / (1 + α (T − 25 °C)) (ISO 7888 practice)
Questions and answers
2500 µS/cm at 45 °C with 2 %/K: what at 25 °C?
1786 µS/cm, TDS about 1161 mg/L at 0.65.
Why 25 °C?
It is the reference for limits and specifications.
Is 2 %/K right for pure water?
No, ultrapure water needs a non-linear correction.
Method and article reviewed by FireCAD engineering on 12 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.