FireCADCalc

Conductivity Temperature Compensation to 25 °C

Conductivity at the 25 °C reference from a reading at another temperature with a linear temperature coefficient, and the TDS by a user factor.

Thermal power plantRefinery and petrochemicalChemical and fertiliser

Inputs

Examples: 2 500 µS/cm at 45 °C, 2 %/KBoiler water 3 800 µS/cm at 25 °CCooling water 900 µS/cm at 12 °C, 1.9 %/K

In µS/cm: the TDS factor is per µS/cm. (Another unit gives κ₂₅ in that unit, but then the TDS is wrong.)
About 0.02 per K (2 %/°C) for most salt solutions; use your sample's value.

Results

Conductivity at 25 °C1,785.7
Ratio of measured to reference1.4000
Estimated TDS (mg/L)1,160.7

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: 2 500 µS/cm at 45 °C, 2 %/K. The inputs, in the units selected above:

InputValue
Measured conductivity2,500.0 -
Sample temperature45.000 °C
Temperature coefficient (per K)0.020000 -
TDS factor (mg/L per µS/cm)0.65000 -

Applying the formula below gives:

ResultValue
Conductivity at 25 °C1,785.7
Ratio of measured to reference1.4000
Estimated TDS (mg/L)1,160.7

Open this example in the calculator and change any input.

Formula

\(\kappa_{25} = \frac{\kappa_T}{1 + \alpha (T - 25\ °C)}, \quad TDS = f\, \kappa_{25}\)

Variables

Symbol Meaning SI unit
\(\kappa_T\) Measured conductivity -
\(T\) Sample temperature K
\(\alpha\) Temperature coefficient (per K) -
\(f\) TDS factor (mg/L per µS/cm) -
\(\kappa_{25}\) Conductivity at 25 °C -
\(\kappa_T/\kappa_{25}\) Ratio of measured to reference -
\(TDS\) Estimated TDS (mg/L) -

Method and standards

Linear compensation.

Where it is used

  • Boiler water and cooling water monitoring.
  • Effluent testing.

Common mistakes

  • Comparing uncompensated readings with limits.
  • One TDS factor for all waters.

Assumptions and limits

  • Linear coefficient over the range. High-purity water (below about 20 µS/cm) and samples more than 25 K from 25 °C are flagged: their temperature dependence is non-linear.
  • Conductivity in µS/cm, the basis of the TDS factor.

References

  • Linear temperature compensation κ₂₅ = κ_T / (1 + α (T − 25 °C)) (ISO 7888 practice)

Questions and answers

2500 µS/cm at 45 °C with 2 %/K: what at 25 °C?

1786 µS/cm, TDS about 1161 mg/L at 0.65.

Why 25 °C?

It is the reference for limits and specifications.

Is 2 %/K right for pure water?

No, ultrapure water needs a non-linear correction.

Method and article reviewed by FireCAD engineering on 12 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.