Chromatography: Theoretical Plates, HETP and Resolution

Number of theoretical plates and plate height of a chromatographic column from a peak's retention time and width, and the resolution between two adjacent peaks.

Thermal power plantProcess steam and utilitiesSugar mill and cogenerationDistillery and ethanolRefinery and petrochemicalChemical and fertiliser

Inputs

Examples: 300 and 325 s on 30 mPacked 2 m column, close peaks

Results

Theoretical plates (peak 2)10,000
Plate height (HETP)3.0000mm
Resolution2.0000

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: 300 and 325 s on 30 m. The inputs, in the units selected above:

InputValue
Retention time, peak 1300.00 s
Base width, peak 112.000 s
Retention time, peak 2325.00 s
Base width, peak 213.000 s
Column length30.000 m

Applying the formula below gives:

ResultValue
Theoretical plates (peak 2)10,000
Plate height (HETP)3.0000 mm
Resolution2.0000

Open this example in the calculator and change any input.

Formula

\(N = 16\left(\frac{t_R}{w}\right)^2, \quad H = \frac LN, \quad R_s = \frac{2(t_{R2} - t_{R1})}{w_1 + w_2}\)

Variables

Symbol Meaning SI unit
\(t_{R1}\) Retention time, peak 1 s
\(w_1\) Base width, peak 1 s
\(t_{R2}\) Retention time, peak 2 s
\(w_2\) Base width, peak 2 s
\(L\) Column length m
\(N\) Theoretical plates (peak 2) -
\(H\) Plate height (HETP) m
\(R_s\) Resolution -

Method and standards

Chromatography definitions.

Where it is used

  • Process GC method checks.
  • Column performance tracking.

Common mistakes

  • Using half-height width with the base-width formula (use 5.54 instead of 16).
  • Mixing time units.

Assumptions and limits

  • Gaussian peaks.

References

  • N = 16(t_R/w)²
  • R_s = 2(t_R2 − t_R1)/(w₁ + w₂) (analytical chemistry definitions)

Questions and answers

What resolution is enough?

1.5 gives baseline separation of similar-sized peaks; process GC methods often target 1.5–2.

How can resolution be improved?

Longer columns (R_s grows with √N), better selectivity (a different phase or temperature) or more retention.

What do the defaults represent?

Two peaks at 300 and 325 s on a 30 m column.

Method and article reviewed by FireCAD engineering on 13 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.