Worked example
Example: 40 m of 4 mm line, 2 L/min gas, 0.2 L filters. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Sample line length | 40.000 m |
| Inside diameter of the line | 4.0000 mm |
| Sample flow (at the analyser, atmospheric) | 2.0000 L/min |
| Mean pressure in the line (absolute) | 1.0132 bar |
| Other volumes (filters, coolers) | 0.20000 L |
| Sample | Gas |
Applying the formula below gives:
| Result | Value |
|---|---|
| Line and component volume | 0.70265 L |
| Transport lag | 21.080 s |
| Velocity in the line | 2.6526 m/s |
Open this example in the calculator and change any input.
Formula
\(V = \frac{\pi d^2}{4} L + V_x, \quad t_d = \frac{V}{Q\,(p_{atm}/p)}\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(L\) | Sample line length | m |
| \(d\) | Inside diameter of the line | m |
| \(Q_N\) | Sample flow (at the analyser, atmospheric) | m³/s |
| \(p\) | Mean pressure in the line (absolute) | Pa |
| \(V_x\) | Other volumes (filters, coolers) | m³ |
| \(\text{fluid}\) | Sample | - |
| \(V\) | Line and component volume | m³ |
| \(t_d\) | Transport lag | s |
| \(v\) | Velocity in the line | m/s |
Method and standards
Residence time of plug flow.
Where it is used
- Combustion and emission analysers.
- Control loops on analyser signals.
Common mistakes
- Forgetting coolers and filters.
- Using line flow at the wrong pressure.
Assumptions and limits
- Plug flow, no mixing.
- The gas flow is corrected for pressure only. In a heated line (for example a 180 °C CEMS line) the gas is also expanded by temperature, so the real velocity is higher and the lag shorter than computed.
References
- Plug-flow residence time t = V / Q (ideal gas at line pressure)
Questions and answers
How long for 40 m of 4 mm line at 2 L/min?
21 s with 0.2 L of filters.
How to cut the lag?
Smaller line, a fast loop or a bypass, and the analyser close to the tap.
Why does pressure matter for gas?
At higher pressure more gas fills the line for the same atmospheric flow.
Method and article reviewed by FireCAD engineering on 12 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.