Worked example
Example: 3 % O₂ at 750 °C against air; 50 mV reading. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Cell temperature | 750.00 °C |
| Oxygen in the sample (by volume) | 3.0000 % |
| Oxygen in the reference gas | 20.950 % |
| Measured cell EMF (to find O₂) | 50.000 mV |
Applying the formula below gives:
| Result | Value |
|---|---|
| Cell EMF for the oxygen entered | 42.839 mV |
| Oxygen for the EMF entered | 2.1679 % |
Open this example in the calculator and change any input.
Formula
\(E = \frac{R\, T}{4 F}\, \ln\frac{O_{2,ref}}{O_2}\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(T\) | Cell temperature | K |
| \(O_2\) | Oxygen in the sample (by volume) | - |
| \(O_{2,ref}\) | Oxygen in the reference gas | - |
| \(E\) | Measured cell EMF (to find O₂) | V |
| \(E_{O_2}\) | Cell EMF for the oxygen entered | V |
| \(O_{2,E}\) | Oxygen for the EMF entered | - |
Method and standards
Nernst equation with four electrons per O₂ molecule.
Where it is used
- Boiler O₂ trim.
- Analyser checks.
Common mistakes
- Mixing wet and dry O₂.
- Cell temperature off its set point.
Assumptions and limits
The default reference, 20.95 % O₂, is dry air. Reference air carrying moisture has slightly less oxygen (about 20.6–20.7 % at typical humidity), which shifts the reading by a few tenths of a millivolt; analysers usually calibrate this out.
Ideal cell, no offset.
References
- Nernst equation for an oxygen concentration cell
- E = (R T / 4F) ln(p_ref / p)
Questions and answers
What EMF does 3 % O₂ give at 750 °C against air?
42.8 mV.
Why does the cell need heating?
Zirconia conducts oxygen ions only when hot, typically above 600 °C.
Why is the scale logarithmic?
The EMF follows the log of the oxygen ratio, so low O₂ gives large signals.
Method and article reviewed by FireCAD engineering on 12 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.