Worked example
Example: 1,413 µS/cm standard reads 1,400 µS; sample 5 kΩ. The inputs, in the units selected above:
| Input | Value |
|---|---|
| Standard solution conductivity at temperature | 1,413.0 µS/cm |
| Measured conductance in the standard | 1,400.0 µS |
| Sample resistance measured | 5,000.0 Ω |
Applying the formula below gives:
| Result | Value |
|---|---|
| Cell constant | 1.0093 1/cm |
| Sample conductance | 200.00 µS |
| Sample conductivity | 201.86 µS/cm |
Open this example in the calculator and change any input.
Formula
\(K = \frac{\kappa_{std}}{G_{std}}, \quad \kappa_s = K\cdot\frac{1}{R_s}\)
Variables
| Symbol | Meaning | SI unit |
|---|---|---|
| \(\kappa_{std}\) | Standard solution conductivity at temperature (µS/cm) | - |
| \(G_{std}\) | Measured conductance in the standard (µS) | - |
| \(R_s\) | Sample resistance measured | Ω |
| \(K\) | Cell constant (1/cm) | - |
| \(G_s\) | Sample conductance (µS) | - |
| \(\kappa_s\) | Sample conductivity (µS/cm) | - |
Method and standards
Definitions.
Where it is used
- Boiler water and condensate analysers.
- Laboratory meter calibration.
Common mistakes
- Calibrating at a different temperature.
- Wrong cell for the range.
Assumptions and limits
- Linear cell; same temperature.
References
- κ = K·G
- K = L/A (electrochemistry definitions; ASTM D1125 principle)
Questions and answers
What is the cell constant?
Electrode spacing divided by area: it converts measured conductance into the liquid's conductivity.
Why different cells for different water?
Ultrapure water has tiny conductance; a small cell constant raises the reading into the meter's best range.
What do the defaults represent?
A cell reading 1,400 µS in a 1,413 µS/cm KCl standard, then 5 kΩ in a sample.
Method and article reviewed by FireCAD engineering on 13 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.