FireCADCalc

Conductivity Cell Constant and Measured Conductivity

Cell constant of a conductivity sensor from its reading in a standard solution, and the conductivity of a sample from its measured conductance or resistance.

Thermal power plantProcess steam and utilitiesSugar mill and cogenerationDistillery and ethanolRefinery and petrochemicalChemical and fertiliser

Inputs

Examples: 1,413 µS/cm standard reads 1,400 µS; sample 5 kΩ0.1 cm⁻¹ cell: 147 µS/cm standard reads 1,520 µS

Results

Standards: 0.01 mol/kg KCl ≈ 1,413 µS/cm at 25 °C. Choose cell constants of 0.01–0.1 cm⁻¹ for pure water and condensate, 1–10 cm⁻¹ for process liquids. Compensate both readings to the same temperature.

Cell constant1.00931/cm
Sample conductance200.00µS
Sample conductivity201.86µS/cm

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: 1,413 µS/cm standard reads 1,400 µS; sample 5 kΩ. The inputs, in the units selected above:

InputValue
Standard solution conductivity at temperature1,413.0 µS/cm
Measured conductance in the standard1,400.0 µS
Sample resistance measured5,000.0 Ω

Applying the formula below gives:

ResultValue
Cell constant1.0093 1/cm
Sample conductance200.00 µS
Sample conductivity201.86 µS/cm

Open this example in the calculator and change any input.

Formula

\(K = \frac{\kappa_{std}}{G_{std}}, \quad \kappa_s = K\cdot\frac{1}{R_s}\)

Variables

Symbol Meaning SI unit
\(\kappa_{std}\) Standard solution conductivity at temperature (µS/cm) -
\(G_{std}\) Measured conductance in the standard (µS) -
\(R_s\) Sample resistance measured Ω
\(K\) Cell constant (1/cm) -
\(G_s\) Sample conductance (µS) -
\(\kappa_s\) Sample conductivity (µS/cm) -

Method and standards

Definitions.

Where it is used

  • Boiler water and condensate analysers.
  • Laboratory meter calibration.

Common mistakes

  • Calibrating at a different temperature.
  • Wrong cell for the range.

Assumptions and limits

  • Linear cell; same temperature.

References

  • κ = K·G
  • K = L/A (electrochemistry definitions; ASTM D1125 principle)

Questions and answers

What is the cell constant?

Electrode spacing divided by area: it converts measured conductance into the liquid's conductivity.

Why different cells for different water?

Ultrapure water has tiny conductance; a small cell constant raises the reading into the meter's best range.

What do the defaults represent?

A cell reading 1,400 µS in a 1,413 µS/cm KCl standard, then 5 kΩ in a sample.

Method and article reviewed by FireCAD engineering on 13 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.