FireCADCalc

Calibration Gas Blending by Partial Pressure

Concentration of a calibration gas blended by partial pressures in a cylinder, and the component fill pressure for a target concentration.

Thermal power plantProcess steam and utilitiesSugar mill and cogenerationDistillery and ethanolRefinery and petrochemicalChemical and fertiliser

Inputs

Examples: 500 ppm from pure gas, 150 bar fill50 ppm from a 1 % parent, 100 bar

Results

Ideal-gas partial pressures; at high pressure, compressibility differences give errors of a few per cent. Gravimetric blending (by mass) is the accurate method for certified standards.

Fill pressure for the target50.000kPa
Concentration achieved with p_c60.000ppm

Download PDF CSV Link to this result Compare another case

Log in to save calculations to your dashboard.

Worked example

Example: 500 ppm from pure gas, 150 bar fill. The inputs, in the units selected above:

InputValue
Final cylinder pressure (absolute)150.00 bar
Component concentration in the source (parent) gas1,000,000 ppm
Target concentration500.00 ppm
Component fill pressure actually used (absolute partial pressure)7.5000 kPa

Applying the formula below gives:

ResultValue
Fill pressure for the target7.5000 kPa
Concentration achieved with p_c500.00 ppm

Open this example in the calculator and change any input.

Formula

\(p_{c,req} = p_t\frac{c_t}{c_s}, \quad c = c_s\frac{p_c}{p_t}\)

Variables

Symbol Meaning SI unit
\(p_t\) Final cylinder pressure (absolute) Pa
\(c_s\) Component concentration in the source (parent) gas -
\(c_t\) Target concentration -
\(p_c\) Component fill pressure actually used (absolute partial pressure) Pa
\(p_{c,req}\) Fill pressure for the target Pa
\(c\) Concentration achieved with p_c -

Method and standards

Dalton's law.

Where it is used

  • Analyser span gas preparation.
  • Detector bump-test mixes.

Common mistakes

  • Using gauge pressures.
  • Ignoring compressibility at high pressure.

Assumptions and limits

  • Ideal gases.
  • Absolute pressures: the partial pressure is the rise from an evacuated cylinder. Gauge readings must have the atmospheric pressure added.

References

  • Dalton's law of partial pressures (definitions)

Questions and answers

Why not just use the source gas?

Analyser spans need concentrations near the measuring range; blending a pure gas or a strong parent mix down saves cost.

How accurate is partial-pressure blending?

A few per cent; certified mixtures are made gravimetrically.

What do the defaults represent?

A 500 ppm mix from a pure component in a cylinder filled to 150 bar.

Method and article reviewed by FireCAD engineering on 13 November 2026. Every result is checked against an independent calculation on each release: how we verify our calculators.

Related calculators

Free engineering calculators: every page computes in SI and converts to your units. Results are for engineering estimation; verify critical designs against the referenced standards.

Rejoining the server...

Rejoin failed... trying again in seconds.

Failed to rejoin.
Please retry or reload the page.

The session has been paused by the server.

Failed to resume the session.
Please retry or reload the page.